Chapter 4 Command Reference
[:SENSe]:FUNCtion Subtree
[:SENSe]:FUNCtion Subtree
This subtree controls the sensor functions.
[:SENSe]:FUNCtion[:OFF] . . . <sensor_function>[,<sensor_function>]
COMMAND
Sets or queries the sensor functions to be sensed by the Counter.
The <sensor_function> strings are:
•
QUERY RESPONSE
•
•
COMMENTS
•
•
Various
RELATED
FRONT-PANEL
KEYS
Programming Guide
"[XNONe:]FREQuency [1 | 2]"
"[XNONe:]POWer [2]"
The query form of this command returns a comma-separated list of
functions that are OFF.
The string omits default nodes (XNONe) and uses short-form
mnemonics. If the channel specifier(s) are set to default value(s), no
channel specifier is returned in response. If the channel specifier(s)
are not set to default value(s), they are returned in the response with
a single space separating the first channel specifier from the function
name.
For example:
–
"FREQ" is returned for frequency on Channel 1.
–
"FREQ 2" is returned for frequency on Channel 2.
*RST: "FREQ 2"
This command can be used to turn individual function(s) OFF without
affecting other functions.
If the optional channel specification is omitted from the
<sensor_function>, a default channel selection is made. For Frequency
and Power, the default is Channel 2.
4-27