Chapter 4 Command Reference
[:SENSe] Subsystem
[:SENSe]:FREQuency:RESolution . . . <numeric_value>[<frequency_unit>]
COMMAND
Sets the resolution of the frequency measurement.
The allowable settings for <numeric_value> and <frequency unit> are:
1 Hz, 10 Hz, 100 Hz, 1 KHz, 10 KHz, 100 KHz, and 1 MHz.
*RST: 1 Hz
COMMENTS
This command does not affect the unit multipliers of any
measurement queries.
If no unit is specified, the global frequency unit in effect is used.
Resol
RELATED
FRONT-PANEL
KEYS
[:SENSe]:FREQuency:TRACking . . . <character_program_data>
COMMAND
Selects one of three signal-tracking modes (SLOW | FAST | OFF). When
TRACking is set to SLOW, the Counter applies two tracking routines after
each measurement; when it is set to FAST, it uses only one tracking
routine; when it is set to OFF, it does not use either of the tracking
routines.
*RST: FAST
COMMENTS
The SLOW setting provides the most accurate signal tracking but
yields the smallest number of measurements in a given period of time.
The FAST setting centers the IF in the IF bandwidth after each
measurement to improve tracking.
None
RELATED
FRONT-PANEL
KEYS
4-26