Menu Structure - Thermo Scientific AquaSensors AquaChlor User Manual

Free chlorine monitor
Table of Contents

Advertisement

5.2. Menu Structure

Pressing the menu key while a measure screen is displayed accesses the main menu. The items in the
main menu are the same regardless of the type of sensor head that is installed in the DataStick. Table 5.1
shows all accessible functions including optional features.
Table 5.1: AV38 Main Menu Functions
Main Menu Option
Calibrate
(Station Number)
Configure
(Station Number)
DataStick Comms
Chlorine Configure
Hold Outputs
Analog Output 1
Analog Output 2 (option)
PID Controller
Relay A (option)
Relay B (option)
Help
Reset AV38
Exit
The main menu is navigated using the up/down arrows to display a desired function. The (
used to indicate main menu choices.
To select a menu choice, press the ENTER key when the desired choice is displayed.
To return to the measure screen, press the ESC key.
When ever there is a value to be entered or a choice to be made, the second line of the display will be
surrounded by parentheses (
the upper line of the screen and be appended with a question mark (
input. Figure 4.1 shows the list of standard menu choices and optional menu choices in the context of the
AV38 display presentation.
Thermo Scientific AquaSensors
Function
Sensor calibration choices are automatically available for installed sensor head.
Enter the MENU from the measure display of interest – station in lower right.
Configuration choices are automatically available for the installed sensor head.
Enter the MENU from the measure display of interest – station in lower right.
Set the station address, baud rate and parity for the Modbus DataStick of
interest.
Select total free chlorine or HOCL concentration for output. Select pH
compensation – manual set or active measure. (Only when Chlorine DataStick
Present).
Holds the 4-20 mA and Relay outputs
Set 4 mA and 20 mA values. Calibrate and test the current output with a meter.
Set 4 mA and 20 mA values. Calibrate and test the current output with a meter.
Set up PID control functions for analog outputs. Available with pH or
Conductivity.
Set timer, alarm, or control functions. (Optional)
Set timer, alarm, or control functions. (Optional)
Indicates AV38 software version
Resets AV38 configuration to default current output, relay and PID settings.
Returns to the measure screen
). The name of the value being entered or choice being made will appear on
( )
AquaChlor
Free Chlorine Monitor User Guide
) to indicate that the user is to provide
?
) symbol is
21