Rack-Mounted Device; Device-Specific Safety Instructions - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

4.4

Rack-mounted device

4.4.1

Device-specific safety instructions

Risk of explosion resulting from improper installation
• Install the devices only in control cabinet/enclosure types that meet the requirements of
• Take suitable ventilation measures to ensure an air exchange amounting to at least 5
• Never introduce occasionally explosive gases!
NOTICE
Insufficient air circulation around the device
Device damage
The device may overheat in case of heat accumulation.
• Make sure that sufficient air circulates around the rack-mounted device.
• Do not cover the ventilation slit.
• If you install more than one rack-mounted device, maintain a clearance of one height
Note
Installation in control cabinets
• Make sure there is adequate ventilation between the devices when installing in control
cabinets.
• Make sure that the permitted ambient temperature listed in the technical specifications is
not exceeded.
Devices in explosion-proof design
Compact Operating Instructions, 05/2018, A5E35134047-03
WARNING
EN/IEC 60079-15 and are correspondingly approved.
times the enclosure volume per hour.
unit (4.5 cm) to the next device.
Installing
4.4 Rack-mounted device
25