Siemens SIPROCESS GA700 Compact Operating Instructions page 90

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
10.2 General technical specifications
Table 10- 19 OXYMAT 7: Influence variables
Influence variables
Ambient temperature
Sample gas pressure
Sample gas flow
Accompanying gases
Supply voltage
Table 10- 20 OXYMAT 7: Climatic conditions
Climatic conditions
Storage and transport
Max. permissible ambient temperature surround-
ing the basic device during operation
Ambient humidity (rel. humidity) during storage,
transport or operation
Restriction for installing an ULTRAMAT 7 analyzer module: 5 to 45 °C
1)
Avoid condensation on the installed components!
2)
90
Deviation at zero point
Deviation of the span
gas
Deviation at zero point
Deviation of the span
gas
Deviation at zero point
Deviation of the span
gas
Analyzer module
24 V DC (-5%/+10%)
Basic device 85 V to
264 V AC / 47 Hz to
63 Hz
1)
2)
≤ 0.5% of the smallest span/10 K or
≤ 50 vpm O
/10 K
2
The larger value applies
≤ 0.5% of the current span/10 K or
≤ 50 vpm O
/10 K
2
The largest value applies
≤ 0.2% of the smallest span/1% pressure change
or
≤ 50 vpm O
/1%, pressure change. The larger
2
value applies
Deviation ≤ 0.2% of current span/1% pressure
change or
50 vpm O
/1% pressure change
2
The larger value applies
≤ 1% of smallest span per 0.1 l/min change in
flow or
50 vpm O
per 0.1 l/min change in flow within the
2
permissible flow range
The larger value applies
≤ 1% of current span per 0.1 l/min change in flow
or
50 vpm O
per 0.1 l/min change in flow within the
2
permissible flow range
The larger value applies
Zero point deviation (cross-sensitivity) in accord-
ance with Table A.1 of EN 61207-3
< 0.1% of the current span
< 0.1% of the current span
- 30 to 70 °C
0 to 50 °C
< 90 %
Devices in explosion-proof design
Compact Operating Instructions, 05/2018, A5E35134047-03