Technical specifications
10.3 Technical specifications for operation in hazardous areas
Pneumatic data of purging gas
Min. purging gas pressure
Consumption during operation at max.
load
Max. load of the Ex p enclosure (continu-
ous load)
Type test pressure on Ex p enclosure
(peak load)
100
For non-combustible sample gases
For combustible sample gases
For Ex py or Ex pz with continuous purging
For Ex pz with simplified pressurized enclo-
sure
< 110 hPa (rel.)
165 hPa (rel.)
Compact Operating Instructions, 05/2018, A5E35134047-03
50 Pa
(rel.)
50 Pa
(rel.)
1 l/min
1 l/min
Devices in explosion-proof design