Dräger Oxylog VE300 Instructions For Use Manual page 95

Emergency and transport ventilator
Hide thumbs Also See for Oxylog VE300:
Table of Contents

Advertisement

Alarm
prior-
ity
!!!
!!!
!!
!
!
!
!
!!
!!
|
Instructions for use
Oxylog VE300 SW 1.n
Alarm
Cause
O
supply
The supply pressure is
2
pressure low
below 1800 mbar
(1800 cmH
Patient dis-
The device has
connected
detected the disconnec-
tion of the breathing cir-
cuit.
Respiratory
The patient breathes at
rate high
a high spontaneous
respiratory rate.
Secondary
Technical fault.
loudspeaker
faulty
Selection not
The selection Standby
confirmed
or Shut down has not
been confirmed with the
rotary knob.
Service date
The scheduled service
reached
date has been missed.
Setting not
The changed setting
confirmed
has not been con-
firmed with the rotary
knob.
VT not
The set tidal volume VT
reached -
was not reached.
leakage
VT not
The set tidal volume
reached -
could not be applied
pressure
because the set pres-
sure limitation was
reached.
Remedy
Make sure that the sup-
ply pressure is above
O).
1800 mbar
2
(1800 cmH
Disconnect the device
from the patient and
continue ventilation
immediately using
another ventilator.
Check the connectors
of the breathing circuit.
Check the patient's
condition, check the
ventilation settings,
adjust the alarm limit if
necessary.
To continue ventilation
with this device, contin-
uously monitor the
device functions. Con-
tact DrägerService.
Check the setting. If
necessary, repeat the
selection and confirm it
with the rotary knob.
Contact DrägerService. 52
Check the setting. If
necessary, adjust and
confirm it with the rotary
knob.
Check the breathing
circuit for leakage.
Check the patient's
condition.
Check the value for
Pmax and adjust as
necessary.
Troubleshooting
Inter-
nal pri-
ority
4
O).
2
11
36
43
47
49
20
37
95