Display installation
Overview possible inputs
1.
DVI
2.
DisplayPort
3.
DVI splitter
3.4 Video output connection
WARNING: When the display is assembled in the medical system, take care of the anchorage of all
cables, to avoid unwanted detachment.
To connect the video outputs
1.
Connect the available display(s) to the corresponding video output(s) using the appropriate video cable(s).
Image 3-3
Overview possible outputs
1.
DVI splitter
The DVI single link connection cables 0.5m included in the box can be used to connect two outputs
of the video splitter to two inputs of the monitor.
3.5 Power supply connection
To connect the power supply
1.
Connect the power cord set into the AC power adapter
2.
Plug the power connector of the adapter into the power port of the display.
Image 3-4
3.
For additional grounding, earth the display by connecting the protective earth pin to a grounded outlet by
means of a wire with a minimum AWG18 size (according to national requirements regarding the maximum
admitted cable length). Use the included M4 screw to attach the wire to the protective earth pin.