Testing Preamp Option (P03, P08, P13, P26); Procedure - Keysight Technologies N9020A MXA Instruction

Signal analyzer
Hide thumbs Also See for N9020A MXA:
Table of Contents

Advertisement

Amplitude Accuracy at 50 MHz

Testing Preamp Option (P03, P08, P13, P26)

Testing Preamp Option (P03, P08, P13, P26)
Instruments containing Options P03, P08, P13, P26 must have the preamp
function turned on and tested.

Procedure

1. On the analyzer, press AMPLTD Y Scale, More, Internal Preamp, Low Band.
2. Connect the Low Power Sensor to the Power Meter, then zero and
3. Connect the equipment as shown in
4. Set the synthesized sweeper amplitude to the value listed in the Nominal
5. Set the signal analyzer input attenuation to 0 dB.
6. Set the Span as listed in
7. Record the source amplitude, as measured by the power meter, in the
8. On the signal analyzer, press Restart.
9. Wait for the analyzer to finish averaging.
10.Press Peak Search.
11.Record the signal amplitude as measured by the analyzer in the measured
12.Calculate the signal amplitude accuracy using the following equation:
13.Record the results under the Amplitude Accuracy Error column of
Table 5-2
Amplitude Accuracy Results (Option P03, P08, P13, P26)
Nominal
Source
Low- band
Amplitude
Preamp
(dBm)
ON
−13
calibrate the meter.
Pad between the adaptor and the signal analyzer input.
Source Amplitude column in
Power Meter Amplitude column of
amplitude column of
Table 5-2
Amplitude Accuracy Error = Meas_amp + 30 dB − Power_meter
5-2
Measured
Res BW
Span
Amplitude
(kHz)
(kHz)
Meas_amp
(dBm)
1
106
Figure 5-1 on page
Table 5-2
Table 5-2
Table 5-2
Power Meter
Amplitude
Power_meter
(dBm)
31
28, using a 30 dB
Table
Amplitude
Accuracy
Test Limit
Error
(dB)
(dB)
±1.13 dB

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents