8. What If ...
27
no
**,***
28
**
32
no
**
33
106
109
110
111
150
531
yes
702
800
801
802
803
1302
**
1501
yes
1510
1522
1541
1800
yes
1815
yes
1820
yes
1825
yes
1830
yes
1835
yes
1900
yes
1901
yes
80
Furnace head cannot be
initialized
The furnace head does not reach
the target position
The vacuum is not released
Necessary vacuum (xxxmbar) is
not reached within 1 min
Timer active – device is in standby The program cannot be started, because the unit was put into stand-by
Incorrect value for HV or V2
HV > H (H2)
Max. number firing protocols
reached
Memory error
Error while trying to store firing
data
Brief power failure during a pro-
gram in progress
Final vacuum value not reached
Vacuum drop
The vacuum does not increase
(self-test)
The vacuum system is not tight
ATK2-calibration:
preheat 962°C
Hint -> working hours
Temperature > VT at the start of a
firing program
SW-update:
Error during copying a file
USB stick
USB stick
USB stick
USB stick
USB stick
Media playback
Data backup – recover
Data backup – recover
The furnace head cannot be brought to the end position. It might be
blocked by external mechanical influences! If this is not the case, contact
your local Ivoclar Vivadent Service Centre.
Furnace head does not open/close correctly. The furnace head is obstructed
or has been moved by hand. Operate the furnace head only with the keys
intended for that purpose.
The vaccum cannot be released. The vacuum valve might be stuck or dirty.
Contact a Service Technician.
The vacuum cannot be built-up. Check the seal of the firing chamber,
vacuum hose, vacuum pump, pump fuse.
mode by the timer (heating turned off). Deactivate the timer or extend the
time frame to be able to execute programs.
Enter plausible values for HV or V2.
Enter a lower value for HV or a higher value for H (H2).
The maximum number of firing program protocol entries has been reached.
A repeated protocol will delete / overwrite existing entries.
Internal memory error. Please restart the device.
An error occurred while logging firing program data. The storage medium
may be full.
A program in progress was interrupted by a short power outage. The
program is being continued.
The set vacuum end value could not be achieved. Check the vacuum
pump.
An unacceptable vacuum drop has occurred.
A vacuum increase could not be measured! Check the following points: Is
the firing chamber tight (no contamination on the sealing surfaces)? Is the
vacuum hose connected? Is the vacuum pump connected? Is the fuse F1 in
order?
The tightness of the vacuum system is not given. Check the sealing
surfaces for possible contamination.
Error during calibration. Sample might be incorrectly inserted. Try again
with a new sample and make sure the sample shows the correct contact.
The heating muffle has ****working hours. Please perform a heating test
(diagnosis) and follow the instructions on the display and in the Operating
Instructions.
The temperature of the firing chamber is higher than the predrying
temperature. To continue the programm nevertheless, press the Start
button.
An error occurred during the software update. Do NOT switch off the
furnace. Try again. If the error reoccurs, try an update via the USB interface.
An error occurred while copying the file. Possible causes: File too large,
invalid data type
No USB memory stick available.
There is no USB stick available or the stick is not prepared as an external
program memory. The USB stick can be prepared in the settings menu.
The USB memory stick is already prepared for programs.
Please unplug all USB devices and then just plug in the USB memory stick
to be prepared.
Not enough free memory.
Playback of the file is not possible.
The new software contains setting options which were not yet available at
the time of the data backup. Only the existing settings of the data backup
were restored.
The new software contains program parameters which were not yet
available at the time of the data backup. Only the existing program
parameters of the data backup were restored.