3 Press Origin to move the Press to its start position.
4 If the lower limit sensor is too high:
5 Press and hold Down to lower the Press until all the probes are
6 If the lower limit sensor is too low:
7 Adjust the position of the lower limit sensor:
i3070 Series 5i Help
Loosen the two screws securing the sensor bracket
move the sensor down to allow the Press to move down.
contacting the board.
Press Slow to slow the movement of the Press if
NOTE
necessary.
Loosen the two screws securing the sensor bracket
a Slide the sensor slowly upwards until it is activated by the sensing
plate (the red sensor light will turn on).
b Slide the sensor down slightly, just enough to turn the sensor light
off. Tighten the two screws on the sensor bracket.
Sensing Plate
Sensor Light
Lower Limit Sensor
Troubleshooting
(Figure 6-
1) and
(Figure 6-
1).