Zone 2 Conveyor Runner Block (Width Adjustment)

Follow this procedure to replace the runner blocks for the Zone 2
conveyor's two horizontal guide rails.
Removal
1 Remove the right side cover.
2 Remove the movable rail assembly:
i3070 Series 5i Help
Movable Rail Assembly
Runner Block (Up/Down)
Runner Block (Width Adjustment)
a Remove six screws securing one side of the rail assembly.
Movable Guide Holder
Runner Block
b Do the same on the other side and remove the rail assembly.
The guide rail can now be accessed.
Vertical Guide Rail
Horizontal Guide Rail
Replacement Procedures
Zone 2 Conveyor Assembly
2 screws
4 screws around
the base
7-19