Table 6-2
Error Messages
Error Message
i3070 Series 5i Help
Cause/Action
The drawer is not in the lowered position.
• Check the drawer's Left Drawer Down Sensor (PHS-23) and/or
Right Drawer Down sensor (PHS-21).
• Lower the drawer.
• Disengage the front safety hook.
• Check the front hook sensor (PHS-17). See
• Disengage the rear safety hook.
• Check the rear hook sensor (PHS-16). See
A board was removed when the system is running in Auto mode.
1 Restart the testplan.
A board is at the exit position of Zone 2 but no status is received from
the controller.
• Remove the board and restart the testplan.
Troubleshooting
Sensor
Troubleshooting.
Sensor
Troubleshooting.
6-7