Environmental Conditions; Electromagnetic Compliance (Emc) Information; Guidance And Manufacturer's Declaration - Invacare Robin Series User Manual

Hide thumbs Also See for Robin Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3
Charger
During charging: 25 – 36 V DC
Voltage output
100 – 240 V AC, 50 – 60 Hz
Voltage supply
Protection class
1
IP41 or IP67 (depending on version)
1
See product label and label on each electric device for
correct protection class. The lowest IP classification decides
the overall classification of the device.
IPx4: Protected against water splashed from any
direction.
IP24: Protected against objects larger than 12.5 mm and
protected against water splashed from any direction.
IP44: Protected against objects larger than 1 mm and
protected against water splashed from any direction.
IP41: Protected against objects larger than 1 mm and
protected against vertically dripping water.
IP67: Dust-tight and protected against temporary
immersion into water.

10.3 Environmental conditions

Storage and
transportation
Temperature
-30 °C to +50 °C
Relative humidity
10 % to 70 %
Atmospheric
pressure
1586745-C
No-load: 41 V DC
Operation
+5 °C to +40 °C
20 % to 90 % at 30
°C, not condensing
700 hPa to 1060 hPa
10.4 Electromagnetic compliance (EMC)
information
Medical Electrical Equipment needs to be installed and used
according to the EMC information in this manual.
This equipment has been tested and found to comply
with EMC limits specified by IEC/EN 60601-1-2 for Class B
equipment.
Portable and mobile RF communications equipment can
affect the operation of this equipment.
Other devices may experience interference from even the
low levels of electromagnetic emissions permitted by the
above standard. To determine if the emission from the lift
is causing the interference, run and stop running the lift.
If the interference with the other device operation stops,
then the lift is causing the interference. In such rare cases,
interference may be reduced or corrected by the following:
Reposition, relocate, or increase the separation between
the devices.

10.4.1 Guidance and manufacturer's declaration

Basic
Test
Standard
Mains Port
EN
Conducted
55011:2007
Emissions
Radiated
EN
Emissions
55011:2007
Technical Data
Compliance Level /
Frequency Range
Group 1 Class B
Group 1 Class B
31

Advertisement

Table of Contents
loading

This manual is also suitable for:

Robin mover

Table of Contents