9 Technical Data

9.1 Characteristics

Automatic Operation
Max. angle in Manual mode 60°
Max. angle in Automatic
mode
Angle values in Automatic
mode
Time values in Automatic
mode

9.2 Weights

Max. user weight
(provided that the weight of the
mattress does not exceed 20 kg)
Max. safe working load weight
(including mattress)
Weights of Soft Tilt components
Complete (main module and 4 arms)
Head arm (1 piece)
Foot arm (1 piece)
Main module
Control Box

9.3 Allowed mattress sizes

Mattress height and side rail compatibility
IMPORTANT
Depending if the end-user in a side position is
laying on a flat wing (mattress horizontal) OR if
the wing the end-user is laying on is raised up to
12°, different combinations of mattresses and side
rails are required (see appropriate chart below).
– End-user laying on a flat wing
Allowed mattress heights, when the end-user is laying on a
flat wing (secondary wing = 0°).
Side rail
Britt V / Line / Line Extended
Britt V / Line / Line Extended
with side rail height extender
mounted*
Scala Medium 2
Scala Decubi 2
*Must be dismounted when the side rail is collapsed and
the end-user is getting out of the bed.
– End-user laying on a elevated wing
Allowed mattress heights, when the end-user is laying on a
up to 12° elevated wing (secondary wing = 1–12°).
1605513-A
yes
primary wing = 30°
secondary wing = 12°
0°–30° (+/- 1° interval)
0, 3, 10–180 min (+/- 10
min intervals)
165 kg
185 kg
25 kg
1.4 kg
2.4 kg
14.5 kg
2.9 kg
Mattress height
12–14 cm
15–29 cm
12–16 cm
17–25 cm
Side rail
Britt V / Line / Line Extended
with side rail height extender
mounted*
Scala Decubi 2
*Must be dismounted when the side rail is collapsed and
the end-user is getting out of the bed.
Mattress width and length depending on bed width
Bed width
Allowed mattress
width
85 cm
83–85 cm
90 cm
88–90 cm
A castellated mattress is recommended.

9.4 Environmental conditions

Storage and
transportation
Temperature
-10°C to +50°C
Relative humidity
Atmospheric
pressure
Be aware that when a bed has been stored under
low temperatures, it must be adjusted to operating
conditions before use.

9.5 Electrical system

Voltage supply: Uin 230 Voltage, AC, 50/60 Hz (AC =
Alternating current)
Maximum current input: Iin max.1,5 Ampere
Intermittent (periodic motor operation): Int = Max. 10
%, 2 min ON / 18 min OFF
Insulation class: CLASS II
Type B Applied Part
Applied Part complying with the specified requirements for
protection against electrical shock according to IEC60601-1.
(An applied parts is a part of the medical equipment
which is designed to come into physical contact with the
patient or parts that are likely to be brought into contact
with the patient.)
Sound level: 58.5 dB (A)
Degree of protection: IPx6* or IPx5**
The main module, actuators and hand control are
protected according to IPx6. The Control box is protected
according to IPx5
Technical Data
Mattress height
12–22 cm
17–18 cm
Min. length (cm)
200 cm
200 cm
Operation
+5°C to +40°C
20% to 75%
800 hPa to 1060 hPa