Technical Data; Characteristics; Weights; Allowed Mattress Sizes - Invacare Medley Ergo Soft Tilt User Manual

Accessory
Table of Contents

Advertisement

9 Technical Data

9.1

Characteristics

Automatic Operation
Max. angle in Manual mode
Max. angle in Automatic mode primary wing = 40°
Angle values in Automatic
mode
Time values in Automatic
mode
9.2

Weights

Max. user weight
(provided that the weight of the mattress
does not exceed 20 kg)
Max. safe working load weight (including
mattress)
Weights of Soft Tilt components
Complete (main module and 4 arms)
Head arm (1 piece)
Foot arm (1 piece)
Main module
Control Box
9.3

Allowed mattress sizes

Mattress height and side rail compatibility
IMPORTANT!
Depending if the end-user in a side position is laying
on a flat wing (mattress horizontal) OR if the wing
the end-user is laying on is raised up to 12°, different
combinations of mattresses and side rails are required
(see appropriate chart below).
- End-user laying on flat wing
Allowed mattress heights, when the end-user is laying on a flat
wing (secondary wing = 0°).
Side rail
Bella /Aria low Mattress support mounted in low
position
Verso ll
Scala Basic 2/ Scala Basic Plus 2
Scala Medium 2
Scala Decubi 2
IVCMAMED221119
yes
80°
secondary wing = 12°
0°–40° (+/- 1° interval)
0, 3, 10–180 min (+/- 10 min
intervals)
116 kg
145 kg
29 kg
3.8 kg
2.6 kg
16 kg
2.5 kg
Mattress
height
12–15 cm
10-12 cm
10-12 cm
10–16 cm
19–25 cm
– End-user laying on a elevated wing
Allowed mattress heights, when the end-user is laying on a up to
12° elevated wing (secondary wing = 1–12°).
Side rail
Scala Medium 2
Scala Decubi 2
Mattress width and length depending on bed
width and side rail
Side rail
Bella /Aria Mattress
support mounted in low
position / Scala Basic2
/ Scala Basic Plus 2/
Scala Medium 2 / Scala
Decubi 2
Verso ll
A castellated mattress is recommended.
9.4

Environmental conditions

Temperature
Relative humidity
Atmospheric pressure
Be aware that when a bed has been stored under
low temperatures, it must be adjusted to operating
conditions before use.
9.5

Electrical system

Voltage supply: Uin 230 Voltage, AC, 50/60 Hz (AC = Alternating
current)
Maximum current input: Iin max.1,5 Ampere
Intermittent (periodic motor operation): Int = Max. 10 %, 2 min
ON / 18 min OFF
Insulation class: CLASS II
Type B Applied Part
Applied Part complying with the specified requirements for pro-
tection against electrical shock according to IEC60601-1.
(An applied parts is a part of the medical equipment which is
designed to come into physical contact with the patient or parts
that are likely to be brought into contact with the patient.)
Matteress
height
Allowed mattress
Min.
width
length
85.5–90 cm
195 cm
85.5-95 cm
Storage and
Operation
transportation
-10°C to +50°C
+5°C to +40°C
20% to 75%
800 hPa to 1060 hPa
Technical Data
10 cm
19 cm
Bed
width
90 cm
21

Advertisement

Table of Contents
loading

This manual is also suitable for:

Soft tilt