Invacare® Octave®

8 Technical Data

8.1 Dimensions bed

All measurements are stated in cm.
All angles are stated in degrees.
All measurements and angles are stated without
tolerances, on a non-extended bed.
81
34
33
52
178
221
57°

8.2 Weights

Max. User Weight (provided that the
weight of the mattress + accessories do
not exceed 35 kg)
Safe Working Load (weight including
mattress + accessories)
Mattress support
Base incl. shear arm
Bed end fitting
Wooden bed end
Complete bed without accessories
Wooden trims (2 pcs)
Wooden side rail Britt V
Lifting pole
Support handle
Mattress (approximately)

8.3 Allowed mattress sizes

Allowed mattress height and length depending on the side
rail in use
Bed side rails
Allowed mattress sizes
Min height
15 cm
Line/Line
Extendable
Britt V
15 cm
16
112
119
131
57°
30°
350 kg
385 kg
72 kg
71 kg
5 kg/pc
8.5 kg/pc
170 kg
2 kg
7.5 kg/pc
10 kg
2 kg
8 kg
Max height
Min length
21 cm
200 cm
21 cm
200 cm
Allowed mattress width (for all side rails)
Bed width
130 cm

8.4 Environmental conditions

Temperature
Relative humidity
Atmospheric
pressure
Allow the product to reach operation temperature
before usage:
Warming up from minimum storage temperature
might take up to 24 hours.
Cooling down from maximum storage
temperature might take up to 24 hours.

8.5 Electrical system

37°
30°
Voltage
Uin 230 Voltage, AC, 50 Hz (AC = Alternating
supply
current)
Maximum
Iin max.3 Ampere
current
input
Intermit-
10 % 2 min ON / 18 min OFF
tent (peri-
odic mo-
tor opera-
tion)
CLASS II equipment
Insulation
class
Type B
Applied Part complying with the specified
Applied
requirements for protection against electrical
Part
shock according to IEC60601-1.
Sound
48 dB (A)
level
Degree of
The control unit, external power supply,
protection
motors and hand controls are protected
according to IPx4 or IPx6 (depending on
configuration). See bed product label and
label on each electric device for correct IP
class. The lowest IP-classification decides the
overall classification of the bed.
IPx4: The system is protected against water
splashed from any direction.
IPx6: The system is protected against water
projected from any direction (not high
pressure).
The bed has no isolator (main switch). If the bed
needs to be electrically disconnected, it has to be
unplugged from the mains socket.
Allowed mattress width
116 – 120 cm
Storage and
Operation
Transportation
-10°C to +50°C
+5°C to +40°C
20% to 90% at 30°C - not condensing
800 hPa to 1060 hPa
1567233-H