RMS-LOG Wireless / LAN Data Loggers
E-M-RMS-LOG-V1.4.docx
Value 1 (Address 31001 / 31002):
Value 2 (Address 31001 / 31002):
0x447A0000
The following website helps for conversion:
https://www.h-
schmidt.net/FloatConverter/IEEE754de.html
NoChange (Big Endian)
31001
16948
0x4234
31002
0
0x0000
31003
17530
0x447A
31004
0
0x0000
Byte Swap (Middle Endian)
31001
13378
0x3442
31002
0
0x0000
31003
31300
0x7A44
31004
0
0x0000
Software Compatibility
The logger is designed for use with the RMS server software (local installation or Rotronic Cloud). The data
logger can alternatively also be operated with the RMS configuration software (standalone software).
© 2018
ROTRONIC AG
Bassersdorf
45.0 = 0x42340000
1000.0 =
Word Swap (Mixed Endian)
31001
31002
31003
31004
Byte and Word Swap (Little Endian)
31001
31002
31003
31004
Switzerland
Instruction Manual
0
0x0000
16948
0x4234
0
0x0000
17530
0x447A
0
0x0000
13378
0x3442
0
0x0000
31300
0x7A44