DSK-1031 INSTRUCTOIN MANUAL DIGITAL BLOOD PRESSURE MONITOR DSK-1031 INSTRUKCJA OBSŁUGI CIŚNIENIOMIERZA CYFROWEGO DSK-1031 РЪКОВОДСТВО ЗА ЕКСПЛОАТАЦИЯ НА ЦИФРОВ ТОНОМЕТЪР DSK-1031 UŽIVATELSKÝ NÁVOD K DIGITÁLNÍMU TONOMETRU DSK-1031 A DSK-1031 TÍPUSÚ DIGITÁLIS TONOMÉTER HASZNÁLATI UTASÍTÁSA MANUAL DE UTILIZARE A TENSIOMETRULUI DIGITAL DSK-1031...
This Instruction Manual is designed to assist the user with safe and eff ective operation of the digital blood pressure monitor DSK-1031 (hereinafter – the “Device”). The device is de- signed to measure systolic and diastolic blood pressure and heart rate reading in patients aged 13 years and older.
Touch control – is control of the device by slight touching it. crease in pressure values. Therefore, blood pressure measured at home is often diff er- ent from that measured in the clinic. Since blood pressure increases at low tempera- tures, measurements should be made at room temperature (about 20°C).
Page 4
ment should be carried out in a quiet comfortable atmosphere at room temperature. POWER SUPPLY OF THE DEVICE Exclude meal an hour before the measurement, and exclude smoking, soft drinks, and alcohol 1.5-2 hours before the measurement. INSTALLATION OF BATTERIES 7.
USE OF THE DEVICE WITH THE POWER SOURCE CORRECT POSITION DURING MEASUREMENT Socket for the power source is arranged on the left side of the device. To use the device Sit at the table, let the chair and with the power source, connect the power source connector to the device, install the table support your back and power source plug into the socket outlet, and press the «START/STOP»...
3. Fix the cuff so that it fi ts tightly to the arm, but see that 4. The symbol « » will disappear, and the measurement will it is not overtight (Fig. 9). Too tight or too free placement start. The cuff pressure will decrease slowly.. of the cuff...
Page 7
8. Press the START/STOP button to turn off the device. Table of Results by the WHO Scale If you forget to turn off the device, it will do so automatically after 3 minutes. Indication WHO Classifi cation Indication of Reliability Hypertension (severe) ≥180 ≥110...
Saved values are kept with the measurement date and time, if set. If you want to save INFORMATION ABOUT ERRORS time and date together with the measurement values, date and time should be set INDICATION LIKELY CAUSE METHODS OF CORRECTION before the measurement.
TECHNICAL SPECIFICATIONS Date and time are Time was not set or measurements Set date and time. displayed as «--/--». were carried out before setting the Date and time can not be saved Operating Principle Oscillometric method time. without setting. Indicator 15 digits liquid crystal display Batteries are discharged.
with a dry, soft cloth. Protection level IP20: Protected against solid foreign particles with a diameter of more than 12.5 mm, no protection against 6. Do not allow the contact between the device and its parts with water, solvents, water. alcohol, and gasoline.
TECHNICAL DESCRIPTION DSK-1031 complies with the Electromagnetic Disturbances standard, IEC60601- 1-2:2014. As a medical electrical equipment, special precautions regarding the electromagnetic disturbances shall be taken at usage of the device according to the information provided below. • The device is not intended for use in environments where the intensity of electromagnetic disturbance is high, such as near active HF surgical equipment and MRI (magnetic resonance imaging) equipment etc.