Table 204. Guidance and manufacturer's declaration - electromagnetic immunity
WSK-1011 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of WSK-1011 should assure that it
is used in such an environment.
Immunity test IEC 60601
test level
Conducted RF
3 Vrms, 150
IEC 61000-4-6
kHz to
80 MHz
Radiated RF
3 V/m, 80
IEC 61000-4-3
MHz to 2.5
GHz
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and refl ection from
structures, objects and people.
1
Field strength from fi xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radios broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fi xed RF transmitters, an electromagnetic site survey should be considered. If the measured fi eld strength in the location in which
WSK-1011 is used exceeds the applicable RF compliance level above, WSK-1011 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating WSK-1011.
2
Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment should be used no closer to any part of WSK-
1011, including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter. Recommended separation distance
3 V
d=[3.5/V1]√P
3 V/m
d=[3.5/E1]√P, 80MHz-800MHz d=[7/E1]√P, 800MHz-2.5GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fi xed RF transmitters, as determined by an electromagnetic site survey,
should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol: