CONFigure Subsystem
CONFigure[:VOLTage][:DC]
Remarks
• You can allow the instrument to automatically select the measurement
• The range is set to autoranging (AUTO) when the <range> parameter
• This command also sets the trigger source to "immediate" and clears all
• Autorange thresholds:
• If the input signal is greater than can be measured on the selected
• If a range change is in progress (due to the autoranging setting) when
82
range using autoranging or you can select a fixed range using manual
ranging. Autoranging is convenient because the instrument decides
which range to use for each measurement based on the input signal.
For faster measurements, use manual ranging on each measurement
(some additional time is required for autoranging since the instrument
has to make a range selection).
is omitted. The resolution is set 5½ digits (MIN) when the
<resolution> parameter is omitted.
calculation functions.
Down range at: <10% of range
Up range at: >120% of range
range (manual ranging), the instrument gives an overload indication.
• Positive overload: "+9.900000E+37" or "OL" from the front panel.
• Negative overload: "-9.900000E+37" or "–OL" from the front panel.
you query the instrument, the value "+9.910000E+37" will be
returned. Wait for the instrument to select an appropriate range before
querying the instrument again.