Cleaning The Probe Base; Cleaning And Disinfection; Accessories; Lifetime - Icare Home Instruction Manual

Hide thumbs Also See for Home:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Icare HOME tonometer Instruction manual for health care professionals

14. CLEANING THE PROBE BASE

You can reuse the probe base after careful cleaning. Clean the probe base
every six months. Change the probe, clean or replace the probe base if the
Service light flashes.
Instructions for cleaning the probe base:
• Fill the probe base cleaning container or other clean container with 70-100%
isopropyl alcohol.
• Turn the power off.
• Unscrew the probe base collar.
• Invert the probe base over the container, drop in the probe base into the
container and let soak for 5-30 minutes.
• Remove the probe base from alcohol.
• Dry the probe base by blowing clean canned or compressed air into the
hole in the probe base. This will additionally remove possible residual dirt.
• Insert the probe base into the tonometer.
• Screw the collar in, to lock the probe base.

15. CLEANING AND DISINFECTION

You must clean the forehead and cheek supports for each new patient. Use
a wipe dampened with a 70-100% isopropyl alcohol solution. Do not immerse
the tonometer in water or other liquid. The tonometer must not be immersed
or cleaned using too much water.

16. ACCESSORIES

Part number Product Description Weight
540
Probe base
TA022-001
Probe base collar
560
Wrist strap
TA022-044
Carrying case
7179
Battery cover
TA022-035
Patient guide
571
Battery 3 V, CR123A
TA022-037
Support position tags 40g
575
USB cable
113
Probe box
543
Probe base cleaning
container

17. LIFETIME

The expected service life of the device is 5 years. The maintenance described
in the chapters 12-15 is required during the the expected service life. The shelf
life of the probes in their intact original packaging is 3 years. We recommend
that you inspect the device for mechanical and functional damage and the safety
labels for legibility annually/every 12 months.
Applicable in Germany only: Messtechnische Kontrolle nach MPG
(Medizinproduktegesetz) alle 24 Monate.

18. TECHNICAL AND PERFORMANCE DATA

Type: TA022
Dimensions: approximately 11cm x 8cm x 3cm.
Weight: approximately 150g.
Power supply: 2 x CR123 non-rechargeable batteries (make sure you use
batteries with built-in PTC protection, for example Energizer Lithium Photo
123 3V CR123A).
8
Dimensions
4g
7 mm x 38 mm
2g
20 mm x 15 mm
3g
270 mm x 10 mm x 10 mm
210g
270 mm x 135 mm x 60 mm
3g
26 mm x 23 mm x 7 mm
33g
210 mm x 90 mm x 2 mm
17g
17 mm x 35 mm
70 mm x 41 mm x 13 mm
23g
1m
55,14g
8,2 x 19,5 x 3,5 cm
3g
5,6 cm x 2 cm

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents