5.5.2 Replace Filters
1. Note orientation of arrow on side of magazine.
2. Insert magazine into filter compartment.
3. Close compartment door
Return to Service
1. Push analyzer into cabinet.
2. Return analyzer to monitor mode in Runtime
Options window.
Section 5 - Maintenance

5.6 Remove and Install Pumps

The Vertex M System includes two vacuum pumps. One
pump operates while the other is idle. You may replace
a defective pump while the other pump continues to
operate.
Note:
You may replace a pump only when the system plac-
es it in standby. Do not replace an operating pump.
To equalize wear on the pumps, alternate the operating
pump every 6 months.
To change operating pump without manually stopping
monitoring touch the Main Screen, Menu, Service,
Authorized Service buttons and then the Pump
Alternate button.
The operating pump indicator is green or yellow; the
indicator for the pump in idle is white.