5.3.1 Remove and Replace Analyzer Par-
ticulate Filters
The Vertex M Chemcassette
filters to protect the unit from particles and potentially
damaging gases. Table 5-1 provides maintenance
information about filters.
The Vertex M Analyzer houses filters in a filter magazine.
Use the following procedure when replacing filters.
1. Set analyzer to idle in the Runtime Options
window.
2. Touch Main Screen, Menu, Maintenance and
then Change Filter.
3. Touch the Analyzer module number in the
Analyzer selection keypad.
4. Touch Release Analyzer.
5. Slide analyzer out of cabinet until filter
compartment is visible.