Vertex M
24-Point Continuous Monitor
TM
4.6.3 Maintenance
Use the Maintenance window to:
•
Load and change Chemcassette
•
Change filters. See
•
Analyzer operation utilities (see the following pages)
Analyzer Operation Window
The Analyzer Operation provides four utility functions which may be useful when performing service on an
analyzer.
•
Tape Advance and Release Analyzer can be performed while analyzer is in monitor.
•
Open Gate and Close Gate can only be performed in idle mode
Open Gate
Opens the optics head to release tape
Section 4 - Operation
tape. See
®
Section 5.3.3 Replace
Section 5.3.4 Change Chemcassette®
Filters.
Close Gate
Closes optics head
4-40
Tape.
Release Analyzer
Unlocks analyzer
so that you may
slide it out of the
cabinet
Advance Tape
Pulls a fresh section
o f t a p e i n t o t h e
d e t e c t i o n h e a d .
Press once to see
stains developed
on points 1-4. Press
twice for points 5-8.