ZOLL M Series Manual page 22

Non-invasive blood pressure
Hide thumbs Also See for M Series:
Table of Contents

Advertisement

N
-I
B
ON
NVASIVE
LOOD
The M Series unit takes an extra measurement without altering the timing of the other measurements.
When the M Series unit is taking automatic measurements, you can stop a single blood pressure
Aborting a
Single
measurement. The M Series unit continues taking all subsequent measurements at the selected
Measurement
intervals. If you press the NIBP button while the M Series unit is not taking a measurement, it takes a
new measurement.
To stop a single measurement:
• Press the NIBP button.
The M Series unit immediately stops taking the measurement and deflates the cuff, but takes all
subsequent automatic measurements at their selected interval.
You can stop all upcoming blood pressure measurements by pressing the NIBP Auto softkey. If the
Stopping all
automatic
M Series unit is taking a measurement, it completes the current measurement but does not take any
measurements
subsequent measurements.
To stop automatic measurements:
• Press the NIBP Auto softkey. If the NIBP Auto softkey label is not displayed, see "Displaying the
NIBP Menu" on page 7.
The M Series unit stops all upcoming measurements.
Aborting Measurements
You can immediately abort a blood pressure measurement at any time by pressing the NIBP button.
You can abort a measurement whether you are in Monitor, Defib or Pacer mode. If you press the NIBP
button while taking a STAT measurement, the M Series unit terminates all measurements.
If you press the NIBP button while the M Series unit is taking an automatic measurement, it aborts the
current measurement, deflates the cuff, and continues all subsequent measurements at the set interval.
See "Taking an Additional Measurement" on page 15 for further details on the use of the NIBP button
while the M Series unit is set to take automatic measurements.
CAUTION
Make sure that a measurement has already started before attempting to abort. If you press the NIBP
button while a measurement is not being taken, the M Series unit starts a new measurement.
To abort the current measurement:
• Press the NIBP button on the M Series front panel.
The M Series unit immediately stops taking the measurement, displays the "NIBP MEAS
ABORTED" message, and deflates the cuff.
16
P
(NIBP)
RESSURE
NIBP button