Printing; Digital Output; Analog Output - Kanomax ANEMOMASTER Series Operation Manual

Table of Contents

Advertisement

13. Troubleshooting

13.4 Printing

Symptom
Printing Failure
Printed characters are too light
or faint.
While printing or feeding paper,
the instrument stops working or
restarts.
Only the value shown on the
display is printed out.
Cannot cancel printing.
"E--1" is displayed on the
screen.
"E--2" is displayed on the
screen.
"E--3" is displayed on the
screen.

13.5 Digital Output

Symptom
Data Transfer Failure

13.6 Analog Output

Symptom
Output Failure
Air Velocity output has a
staircase pattern.
Incorrect Output Value
Possible Cause(s) / Solution(s)
Printing paper may not be installed properly.
→ Open the printer cover and confirm there is no ejected paper
caught inside the printer.
Printing paper may be running short.
→ If a red mark appears on the edges of the paper, install a new
printing paper.
The batteries may be low.
Turn the power OFF and replace the batteries.
The batteries may be low.
Turn the power OFF and replace the batteries.
The batteries may be low.
→ Turn the power OFF and replace the batteries.
→ If you print data often or print out a large amount of data, use
the AC adaptor (sold separately).
Single parameter output mode is activated in the data output
setting.
Switch to all parameters output mode in the function setting.
You cannot cancel printing.
The printer paper is not set.
Check to ensure printer paper is properly set in the roller.
The battery is low.
Turn the power OFF and replace the batteries.
Printer Head has overheated due to continuous printer use.
Stop printing temporarily and allow printer to cool.
The Printer Head Lever is up.
Pull the lever down.
Printer Head Lever
Possible Cause(s) / Solution(s)
The RS232C cable may not be connected properly.
Communication cable exclusively for this model is required.
The baud rate setting may not be correct.
Check the settings of the ANEMOMASTER and the computer.
Confirm that the communication command is correct.
Possible Cause(s) / Solution(s)
Confirm that the polarity of the output terminal is correct.
Confirm that normal measurement mode is selected.
Check the Time Constant (TC) setting.
Confirm that the analog output setting is correct.
Confirm that the output range setting is correct.
The load impedance may be set lower than the defined value.
(Load impedance: 5kΩ or higher)
38
Refer To
(Page No.)
19
19
6
6
23
29
19
6
19
Refer To
(Page No.)
24
24
25
Refer To
(Page No.)
27
11
13
27
27
27