Entering Command From Pc To Output Data; Transmission Of Raw Data (Data Measured Every 1Sec) - Kanomax ANEMOMASTER Series Operation Manual

Table of Contents

Advertisement

6. Data Output

6.3 Entering Command from PC to Output Data

To connect the ANEMOMASTER to your PC, refer to "6.2.1 Preparation for Digital Output" in page 24.
---Icon and its Meaning---
: Space
: Line Break or Press Enter
*: To enter number
* Enter all commands in capitals.

6.3.1 Transmission of Raw Data (data measured every 1sec)

DISPLAY
Example:
When entering "D0005" in the Air
Velocity Measurement Mode.
AD
0.19;
26.8;
0.51;
26.8;
0.61;
27.0;
0.24;
27.0;
0.15;
26.9;
AN
DISPLAY
AS
VT-;01;
DISPLAY
AU
m/s;℃;kPa
<To set how many data to read>
Enter "D**** ". (Enter the number of data to be read in 4 digit number.)
After the command is received, "AD" will be returned. Then the raw data
that appears on the instrument's screen and is updated every second will be
73.4
output. The maximum amount of data which can be set is 9999 data. To
read more than 9999 data, send another command.
73.5
76.1
<Output >
77.5
With Pressure Measurement Function (Optional):
Air Velocity; Air Temperature; Pressure
76.0
Without Pressure Measurement Function:
Air Velocity; Air Temperature; 0000000
<Receive Interrupt>
Enter "N ". After the command is received, "AN" will be returned and the
reception will be interrupted.
<To output measurement condition>
Enter "S ". After the command is received, "AS" will be returned. The
measurement parameters displayed on the screen and time constant (See
page 13) will be output.
<Output>
Measuring Parameter (VT-, PRS); Time Constant;
<To output measurement unit>
Enter "U ". After the command is received, "AU" will be returned.
Configured measurement at the time will be output.
<Output Content>
Air Velocity Unit; Temperature Unit; Pressure Unit
Command
D****
To set how many data to read
N
Receive Interrupt
S
To output measurement condition
U
To output measurement unit
P
To output stored data volume.
T****
To output stored data
PROCEDURE
PROCEDURE
VT-: Air Velocity or Air Temperature
PRS: Pressure (Optional)
PROCEDURE
Function
25