6. Data Output

6.2 Digital Output

6.2.1 Preparation for Digital Output

To download the data stored in the ANEMOMASTER to your PC, connect the
ANEMOMASTER to your PC using the RS232C cable (Optional).
<What you need>
●Personal Computer
●RS-232C Cable (Sold Separately)
●Communication Software (Example: Windows has hyper terminal software.)
<Baud Rate Setting>
The baud rate setting of the instrument and the PC must be the same.
The baud rate setting of the main unit is as follows:
Data Bit Length
Parity
Stop Bit
Delimiter
Baud Rate
*As for the baud rate setting, refer to "7.2 Other Function Settings" in page 29.
As for your PC setting, refer to the operation manual of your PC
<Connecting the instrument to a PC>
(1) Connect the instrument to a PC by inserting a connecting cable into RS232C terminal on the instrument.
(2) Turn ON the instrument.
Make sure that instrument is in Normal Measurement mode.
RS232C Cable Wiring Diagram
PC (D-Sub9 pin)
Signal
Pin No.
NC
RXD
TXD
NC
GND
NC
RTS
CTS
NC
* The measurement software is available for Windows. (Sold separately)
8 bits
None
1
CRLF
Based on the setting value*
Connection
Pin No.
1
2
3
4
5
6
7
8
9
ANEMOMASTER
Signal
Description of Signal
1
GND
2
TXD
Transmitting Data
3
RXD
4
CTS
5
RTS
6
NC
RS232C Terminal
Signal Ground
Receive Data
Clear to Send
Request to Send
24
Signal Direction
Output
Input
Input
Output