TM-2
..
--
Table B-1.
Command and Monitor Data and Format (Cont.)
The command:
1
2
'3
4
5
6
7
Filter Modification
Lower Sideband Filter
Upper Sideband
1.
7 kHz Symmetric:
400 Hz symmetric
Unused
3.2 kHz Symmetric
Unchanged from previous setting
Command 8 - Enable AGC Dump
This command causes all subsequent commands containing receiver data
to cause an automatic AGC dump.
Command 9 - Inhibit AGC Dump
B-8
This command causes the AGC command dump to be disabled.
This would
be used when free ,tuning the receiver by remote control.
2)
Monitor Status
The ---following summary sta tus is apPended to all other moni tor
data streams:
1 - Receiver operating in remote control
2 - Synthesizer out of lock
4 - Receiver in Override remote control mode
8 - Last command sequence had character transmission
erro~
16 - Last command
seq~ence
had data error
32 - Lost data error in last sequence
The indicators above are added together to from the summary
status.
For instance 17
indic~tes
remote operation with a data
error in the last command stream.

Advertisement