Figure 22: Fluidics Components
Thermo Scientific
When rinsing the flow cell before a low level standard calibration, be sure to rinse off the
inside of the reservoir lid with deionized water to prevent build up of sodium contamination
along the lid.
To ensure the analyzer achieves stability and minimize interference during calibration, the
operator should minimize activity on system during calibration.
Rinsing the Flow Cell
WARNING: Be sure to follow the procedure for rinsing the flow cell.
1. Open the sample inlet valve, see
Position the valve with the pointed end facing up for open position.
2. Open the bypass/needle valve by turning it counterclockwise, see
Check for the appropriate sample flow.
Adjust the pressure regulator if required. See step 3.
A
4
|
Calibration
Section
B
.
Orion 2111XP Sodium Analyzer
|
57