Use Proper SMA Connector Care and Connection Techniques
3 Hold the torque wrench lightly, at the end of the handle only (beyond the
Figure 9
4 Apply force downward to the wrench handle. This applies torque to the
5 Tighten the connection just to the torque wrench break point. The wrench

Separating Connections

To avoid lateral (bending) force on the connector mating plane surfaces, always
support the devices and connections.
1 Use an open-end wrench or spanner wrench to prevent the device body from
2 Use another open-end wrench or the torque wrench to loosen the connector
3 Complete the separation by hand, turning only the connector nut.
4 Pull the connectors straight apart without twisting, rocking, or bending.
Keysight PD1000A Power Device Measurement System
groove). See the following figure.
Torquing Direction
Stop when handle begins to yield
Using the Torque Wrench
connection through the wrench. Do not hold the wrench so tightly that you
push the handle straight down along its length rather than pivoting it,
otherwise you apply an unknown amount of torque.
You don't have to fully break the handle of the torque wrench to
reach the specified torque; doing so can cause the handle to kick
back and loosen the connection. Any give at all in the handle is
sufficient torque.
handle gives way at its internal pivot point. Do not tighten the connection
further.
Do
turn the device body. Only turn the connector nut. Damage
not
to the center conductor can occur if the device body is twisted.
turning.
nut.
23