Test Fixture Calibration
11 Insert the Port 1 SHORT Cal Standard into the test fixture. The SHORT Cal
12 Insert the Port 1 50 Ω LOAD Cal Standard into the test fixture. The 50 Ω LOAD
13 Insert the Port 2 OPEN Cal Standard into the test fixture. The Port 2 OPEN Cal
14 Insert the Port 2 SHORT Cal Standard into the test fixture. The SHORT Cal
Keysight PD1000A Test Fixture Operating Guide
Standard goes between the Gate and Drain/Collector. Click Measure.
Cal Standard goes between the Gate and Drain/Collector. Click Measure.
Standard goes between the Source/Emitter and Drain/Collector. Click
Measure. Note: for the TO-220/TO-247 Test Fixtures, use the same OPEN Cal.
Standard as before, but turn it around to fit into the correct sockets on the
test fixture.
Standard goes between the Source/Emitter and Drain/Collector. Click
Measure. Note: for the TO-220/TO-247 Test Fixtures, use the same SHORT Cal.
Standard as before, but turn it around to fit into the correct sockets on the
test fixture.
25