Disassembly procedure
Disassemble the power sensor by performing the following steps:
1 Disassemble the power sensor only in a static free workstation. Electrostatic
Figure 3-8
2 At the rear of the power sensor, insert the blade of a screwdriver between the
3 Pry alternately at both sides of the connector J1 until the plastic shells are
Reassembly procedure
Replace the magnetic shields and the plastic shells. Snap the plastic shells
together.
Keysight E9320 Operating and Service Guide
discharge renders the power sensor inoperative.
Removing the power sensor shell
plastic shells (See
Figure
screwdriver blade as wide as the slot between the two shells.
apart. Remove the shells and the magnetic shields.
3-8). To prevent damage to the plastic shells use a
Service
3
61