Settling Times; Table 2-15 Averages Vs. Settling Time (Average-Only Mode) - Keysight Technologies E9320 Series Operating And Service Manual

Peak and average power sensors
Hide thumbs Also See for E9320 Series:
Table of Contents

Advertisement

Settling times

Average-only mode
In normal and x2 speed, manual filter, 10 dB decreasing power step, refer to
Table
Table 2-15
In fast mode, within the range –50 to +20 dBm, for a 10 dB decreasing power
step, the settling time is:
E4416A: 10 ms
E4417A: 20 ms
When a power step crosses the power sensor's auto-range switch point, add
25 ms.
Keysight E9320 Operating and Service Guide
2-15.
Averages vs. settling time (average-only mode)
Number of averages
1
2
4
8
16
32
64
128
256
512
1024
Specifications and Characteristics
Settling time (s)
Normal speed
0.08
0.13
0.24
0.45
1.1
1.9
3.5
6.7
14
27
57
2
x2 speed
0.07
0.09
0.15
0.24
0.45
1.1
1.9
3.5
6.7
14
27
39