Clean the connector only at a static free workstation. Electrostatic discharge to the
C A U T I O N
center pin of the connector will render the power sensor inoperative.
Clean the connector face using a cotton swab dipped in isopropyl alcohol.
If the swab is too big use a round wooden toothpick wrapped in a lint
free cotton cloth dipped in isopropyl alcohol. Refer to Keysight Application
Note 326, Principals of Microwave Connector Care (5954- 1566) or
Microwave Connector Care (08510- 90064) for proper cleaning methods.

Disassembly Procedure

Figure 5
Diassemble the power sensor only in a static free workstation. Electrostatic discharge
C A U T I O N
will render the power sensor inoperative.
22
Removing Power Sensor Shell
E4412A and E4413A Operating and Service Guide