General Information
This chapter contains information about the general maintenance, performance
tests, troubleshooting, and repair of E-Series E9320 power sensors.
Cleaning
Use a clean, damp cloth to clean the body of the E-Series E9320 power sensor.
Connector cleaning
The RF connector beads deteriorate when contacted by hydrocarbon compounds
such as acetone, trichloroethylene, carbon tetrachloride, and benzene.
Clean the connector only at a static free workstation. Electrostatic discharge to
the center pin of the connector will render the power sensor inoperative.
Keeping in mind its flammable nature; a solution of pure isopropyl or ethyl alcohol
can be used to clean the connector.
Clean the connector face using a cotton swab dipped in isopropyl alcohol. If the
swab is too big use a round wooden toothpick wrapped in a lint free cotton cloth
dipped in isopropyl alcohol.
44
Keysight E9320 Operating and Service Guide