Keysight Technologies E9320 Series Operating And Service Manual page 47

Peak and average power sensors
Hide thumbs Also See for E9320 Series:
Table of Contents

Advertisement

Function generator (33611A)
Signal generator (N5182B)
Power meter (E4416/7A)
E-Series E9320 power sensor
Sensor Cable (E9288A)
DUT
Figure 3-3
4 Perform sensor zero and calibration and connect the sensor to the signal
5 Set the following for signal generator (N5182B):
– Turn on IQ state
– Set IQ modulator source to EXTERNAL
– Set frequency to 1 GHz
– Set power level to 0 dBm
6 Set the following for 33611A:
– Output a PULSE signal with frequency 10 kHz
– Set amplitude to 0.5Vpp and offset to 0.25V
– Set pulse width to 50us
Keysight E9320 Operating and Service Guide
Rise and fall time performance test equipment setup
generator (N5182B) RF output port once completed.
OUTPUT
SYNC OUT
I INPUT
RF OUT
RF IN
TRIG IN
Service
3