Table 2-4
[a] Applies to CW and constant amplitude signals only above –20 dBm.
Table 2-5
[a] Applies to CW and constant amplitude signals only above –20 dBm.
Keysight E9320 Operating and Service Guide
E9322A and E9326A lower and upper measurement ranges
Lower range (min. power)
Lower range (max. power)
Lower to upper auto range point
Upper to lower auto range point
Upper range (min. power)
Upper range (max. power)
E9323A and E9327A lower and upper measurement ranges
Lower range (min. power)
Lower range (max. power)
Lower to upper auto range point
Upper to lower auto range point
Upper range (min. power)
Upper range (max. power)
Specifications and Characteristics
E9322A and E9326A
Normal mode
–45 dBm
–5 dBm
–15 dBm
–35 dBm
+20 dBm
E9323A and E9327A
Normal mode
–40 dBm
–5 dBm
–15 dBm
–30 dBm
+20 dBm
2
Average-onl y mode
–60 dBm
[a]
–13.5 dBm
–14.5 dBm
–45 dBm
[a]
+20 dBm
Average-onl y mode
–60 dBm
[a]
–10.5 dBm
–11.5 dBm
–35 dBm
[a]
+20 dBm
25