Keysight E4412A and
E4413A Power Sensors
Operating and
Service Guide

Advertisement

Table of Contents
loading

Summary of Contents for Keysight Technologies E4412A

  • Page 1 Keysight E4412A and E4413A Power Sensors Operating and Service Guide...
  • Page 2: Safety Information

    Do not proceed beyond a WARNING acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR notice until the indicated conditions are 227.7103-5 (c), as applicable in any fully understood and met. technical data. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 3: Certification

    Buyer shall prepay shipping charges to Keysight and Keysight shall pay shipping charges to return the product to the Buyer. However, the Buyer shall pay all shipping charges, duties, and taxes for products returned to Keysight from another country. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 4: Limitation Of Warranty

    Keysight shall not be liable for any direct, indirect, special, incidental, or consequential damages (including lost profit or data), whether based on warranty, contract, tort, or any other legal theory. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 5: Safety Summary

    Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies, Inc. assumes no liability for the customer’s failure to comply with these requirements.
  • Page 6: Safety Symbols

    On (Supply). Off (Supply). Equipment protected throughout by DOUBLE INSULATION or REINFORCED Caution, risk of electric shock. INSULATION. Caution, hot surface. In position of bi-stable push control. Out position of bi-stable push control. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 7: Environmental Conditions

    Environmental Conditions The E4412A and E4413A is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument. Environmental cond ition Requirement Operating condition – 0 to 55 ºC...
  • Page 8: Regulatory Markings

    This ISM device complies with the equipment. The affixed product label Canadian ICES-001. indicates that you must not discard Cet appareil ISM est conforme à la this electrical/electronic product in norme NMB-001 du Canada. domestic household waste. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 9: General Safety Information

    – Notice for germany: Noise Declaration L ρ A < 70 dBm am Arbeitsplatz (operator position) mormaler Betrieb (normal position) nach DIN 45635 T.19 (per ISO 7779). Keysight E4412A and E4413A Operating and Service Guide...
  • Page 10: Waste Electrical And Electronic Equipment (Weee) Directivel

    To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites: – www.keysight.com/find/powersensors (product-specific information and support, software and documentation updates) – www.keysight.com/find/assist (worldwide contact information for repair and service) Keysight E4412A and E4413A Operating and Service Guide...
  • Page 11: Table Of Contents

    ........29 Keysight E4412A and E4413A Operating and Service Guide...
  • Page 12 ..........41 Appendix Keysight E4412A and E4413A Operating and Service Guide...
  • Page 13: Keysight E4412A And E4413A Operating And Service Guide

    ....34 Figure 1-8 Removing power sensor shell ....40 Keysight E4412A and E4413A Operating and Service Guide...
  • Page 14 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 15: List Of Tables

    US3848/MYxxxxxxxx/SGxxxxxxxx) ....44 Table A-2 E4412A calibration factor uncertainty at 1 mW (0 dBm) (serial prefixes below US3848/MYxxxxxxxx/ SGxxxxxxxx) .......46...
  • Page 16 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 17 Keysight E4412A and E4413A Power Sensors Operating and Service Guide Operation and Service Guide General Information Performance Test Replaceable Parts Service Adjustments This chapter contains information about initial inspection, performance tests, specifications, operations, troubleshooting and service of the Keysight E4412A and E4413A power sensors.
  • Page 18 Operation and Service Guide Figure 1-1 E4412A and E4413A power sensors (formerly ECP-E18A and EXCP-E26A, respectively) Keysight E4412A and E4413A Operating and Service Guide...
  • Page 19: Operation And Service Guide

    CW microwave power levels in a wide dynamic range from -70 dBm to +20 dBm (100 pW to 100 mW). The E4412A measures at frequencies from 10 MHz to 18.0 GHz. The E4413A measures at frequencies from 50 MHz to 26.5 GHz.
  • Page 20: Specifications

    The E4412A and E4413A power sensors are extremely static- sensitive. Do not NOTE open the power sensor unless you and the power sensor are at a static-free workstation. Specifications For the specifications of the E4412A and E4413A power sensors, refer to the datasheet at http://literature.cdn.keysight.com/litweb/pdf/5990-4019EN.pdf. ±6 % ±3 % –20 –35...
  • Page 21 Reference power = –25 dBm; power measured = –35 dBm, for the case of an insertion loss of 10 dB. the chart indicates a typical error of ±1 %, which corresponds to ±0.04 dBm uncertainty in the measurement. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 22: Calibration Factor (Cf) And Reflection Coefficient (Rho)

    30 MHz to <2 GHz 2 GHz to <16 GHz 16 GHz to 18 GHz [a] Specifications apply to instruments with serial prefix US3848/MYxxxxxxxx/SGxxxxxxxx and above. For earlier instruments, refer to Appendix A. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 23: Installation

    If there is mechanical damage or if the instrument does not pass the performance tests, notify the nearest Keysight Technologies office. Keep the damaged shipping materials (if any) for inspection by the carrier and a Keysight Technologies representative. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 24: Interconnections

    Operation and Service Guide Interconnections Connect one end of the 11730A sensor cable to the E4412A or E4413A power sensor and connect the other end of the cable to the power meter’s channel input. Allow a few seconds for the power meter to download the power sensor’s calibration table before making a measurement.
  • Page 25: Operation

    The connector plastic insulator bead deteriorates when contacted by acetone, trichloroethylene, carbon tetrachloride, benzene, etc. Refer to Application Note 326, Principals of Microwave Connector Care (5954-1566) or Microwave Connector Care (08510-90064) for proper cleaning methods. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 26: Power Meter Calibrations

    – The adapter supplied with a new sensor typically has a loss of < 0.1%. Adapter loss could potentially contribute to absolute power measurement accuracy and needs to be considered. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 27: Adapter Test Procedure

    S12 and S21 0.999235 dB (lin mag) 0.999205 dB (lin mag) <0.1% <0.1% Open, Short, Load, Sliding load Port 1 Open, Short, Load, Sliding load Port 2 Figure 1-4 System calibration setup Keysight E4412A and E4413A Operating and Service Guide...
  • Page 28 9 The firmware will estimate the delay, which should be approximately 0.11 ns. Device-Under-Test (DUT) Measurement PNA Port 1 PNA Port 2 Figure 1-5 DUT measurement setup 1 Leave the adapter-under-test in place - refer to step Keysight E4412A and E4413A Operating and Service Guide...
  • Page 29: Operating Instructions

    Operating instructions The E4412A and E4413A power sensors are compatible ONLY with the newer E44XX-Series power meters. They are NOT compatible with the earlier 430-Series, E1416A, or 70100A power meters. To operate the power sensor, refer to the operating instructions in the Keysight E44XX-Series Power Meter User’s...
  • Page 30: Performance Test

    Table 1-5 when testing the E4412A, or in Table 1-6 when testing the E4413A. Table 1-5 Power sensor SWR and reflection coefficient for the E4412A Frequency System Rho uncertainty Actual measurement Maximum Rho 10 MHz to <30 MHz ± 0.010 0.099...
  • Page 31: Zero Set Performance Verification

    Ideally, this difference should be zero. This performance verification requires the E4416/7A power meter. Procedure 1 Connect the DUT (E-Series E4412A or E4413A power sensor) to the power meter as shown in Figure 1-6.
  • Page 32 11 step 12 and then calculate the mean value of the readings. 13 Compare the calculated mean value to the product datasheet. If the verification fails, refer to “Adjustments” on page 41. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 33: Replaceable Parts

    Roseville, California. Ask your nearest Keysight office for information and forms for the “Direct Mail Order System.” Also your nearest Keysight office can supply toll free telephone numbers for ordering parts and supplies. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 34: Figure 1-7 Illustrated Parts Break Down

    Operation and Service Guide Figure 1-7 Illustrated parts break down Keysight E4412A and E4413A Operating and Service Guide...
  • Page 35: Replaceable Parts

    SHELL PLASTIC 08481-20011 CHASSIS 08481-20011 CHASSIS 08481-00002 SHIELD 08481-00002 SHIELD MP26 E4412-80002 LABEL, ID E4412A MP26 E4413-80002 LABEL, ID E4413A MP27 7121-7389 LABEL, POWER SENSOR MP30 7121-7388 LABEL, CAL/ESD MP31 00346-80011 LABEL, CAUTION Keysight E4412A and E4413A Operating and Service Guide...
  • Page 36: Service

    EEPROM. This configures the power meter to operate over the +20 dBm power range with that particular sensor’s unique correction data applied. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 37: Troubleshooting

    Subjecting a power sensor module above its maximum allowable power rating is considered a misuse or self-abuse and is excluded from Keysight Technologies’ standard warranty coverage. Refer to the Power Sensor Overpower Failure Verification Guideline at https://literature.cdn.keysight.com/litweb/pdf/5992-4039EN.pdf...
  • Page 38: Connector Maintenance

    – Do not overtighten connectors and do not allow them to become loose — use a torque wrench. – Do not mate dissimilar families, for example APC-3.5 and SMA. – Avoid temperature extremes. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 39: Cleaning

    Refer to Keysight Application Note 326, Principals of Microwave Connector Care (5954-1566) or Microwave Connector Care (08510-90064) for proper cleaning methods. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 40: Disassembly Procedure

    2 Pry alternately at both sides of the connector J1 until the plastic shells are apart. Remove the shells and the magnetic shields. Reassembly procedure 1 Replace the magnetic shields and the plastic shells as shown in Figure 1-8. Snap the plastic shells together. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 41: Adjustments

    Performance verification must be completed after any repairs that may have altered the characteristics of the E-Series E4412A and E4413A power sensors. The E-Series E4412A and E4413A power sensors can be returned to Keysight for adjustments.
  • Page 42 Operation and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 43: E4412A And E4413A Specifications

    Operating and Service Guide Appendix E4412A and E4413A specifications (serial prefixes below US3848/MYxxxxxxxx/ SGxxxxxxxx) E4412A calibration factor uncertainty at 1 mW (0 dBm) (serial prefixes below US3848/MYxxxxxxxx/SGxxxxxxxx) E4413A calibration factor uncertainty at 1 mW (0 dBm) (serial prefixes below US3848/MYxxxxxxxx/SGxxxxxxxx)
  • Page 44 Appendix Table A-1 E4412A and E4413A specifications (serial prefixes below US3848/MYxxxxxxxx/ SGxxxxxxxx) Limit Comments Frequency Range E4412A: 10 MHz to 18 GHz E4413A: 50 MHz to 26.5 GHz Power Range –70 dBm to +20 dBm (100 pW to 100 mW)
  • Page 45 Appendix Table A-1 E4412A and E4413A specifications (serial prefixes below US3848/MYxxxxxxxx/ SGxxxxxxxx) (continued) Limit Comments Net Weight E4412A 0.47 kg (1.04 lb) E4413A 0.45 kg (1.00 lb) Dimensions E4412A Length: 130 mm (5.1 in) E4413A Length: 102 mm (4.0 in) Wid th: 38 mm (1.5 in)
  • Page 46: Us3848/Myxxxxxxxx/Sgxxxxxxxx)

    Appendix Table A-2 E4412A calibration factor uncertainty at 1 mW (0 dBm) (serial prefixes below US3848/MYxxxxxxxx/SGxxxxxxxx) Frequency Uncertainty (%) 10 MHz 30 MHz 50 MHz Reference 100 MHz 1.0 GHz 2.0 GHz 4.0 GHz 6.0 GHz 8.0 GHz 10.0 GHz 11.0 GHz...
  • Page 47: E4413A Calibration Factor Uncertainty At 1 Mw (0 Dbm) (Serial Prefixes Below Us3848/Myxxxxxxxx/Sgxxxxxxxx)

    1.0 GHz 2.0 GHz 4.0 GHz 6.0 GHz 8.0 GHz 10.0 GHz 12.0 GHz 14.0 GHz 16.0 GHz 17.0 GHz 18.0 GHz 20.0 GHz 22.0 GHz 24.0 GHz 26.0 GHz 26.5 GHz Keysight E4412A and E4413A Operating and Service Guide...
  • Page 48 Appendix THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight E4412A and E4413A Operating and Service Guide...
  • Page 49 This information is subject to change without notice. Always refer to the Keysight website for the latest revision. © Keysight Technologies 1999 - 2019 Edition 9, September 4, 2019 Printed in Malaysia *E4412-90013* E4412-90013 www.keysight.com...