Warnings and cautions
WARNING: Electrical shock
Do not connect the ECG device to an electrical outlet controlled
by a wall switch or dimmer.
WARNING: Electrical shock
To reduce the risk of electrical shock, do not remove the cover (or
back) of the ECG device. Refer servicing to qualified personnel.
WARNING: Electrical shock
To reduce the risk of shock, always disconnect the device.from
power before cleaning.
WARNING: Potential for electric shock
Cables in a computer network that lead outward from a system
are typically electrically conducting connections. Do not connect
any non-isolated accessories to this device.
WARNING! Potential for electric shock
If any matter (gaseous, liquid, or solid matter) enters the ECG
device enclosure, you must clean the device and then have it
inspected by Cardiac Science service personnel.
WARNING! Potential for electric shock
To prevent current from a non-medical electronic device
(computer, monitor, printer) from reaching the patient, the
operator must not touch the non-medical device and the patient
at the same time.
WARNING! Potential for electric shock
The ECG device was not designed to support USB devices that
connect to a standard wall socket. Do not connect externally
powered USB devices to this equipment.
WARNING! Fire, explosion, or contamination
Batteries must be disposed of properly. Dispose of batteries in
accordance with local regulations.