Cardiac Science Burdick 8500 Product Safety Manual page 11

Electrocardiograph
Hide thumbs Also See for Burdick 8500:
Table of Contents

Advertisement

Warnings and cautions
WARNING! False readings
The lead status check does not detect lead reversals.
Always ensure the correct lead placement is used for the lead
configuration selected.
WARNING! Misdiagnosis
Dirty, compromised, or unprepared skin may cause incorrect
readings.
Before applying electrodes, ensure the skin is clean, unbroken,
and properly prepared.
WARNING! Infection
Lead placement on compromised skin may cause infection.
Before applying electrodes, ensure the skin is clean, unbroken and
properly prepared.
Caution: Patient skin irritation
When applying electrodes, skin preparation, electrolyte solution,
or electrode material may cause skin irritation. Monitor the
electrode site and, if irritation occurs, use alternative electrolyte or
electrodes.
Caution: Patient safety
Disposable electrodes are single use only. Never re-use disposable
electrodes.
Caution: Baseline drift
Never mix patient electrode types or brands. Dissimilar metals or
other incompatibilities may cause considerable baseline drift and
may increase trace recovery time after defibrillation.
Defibrillation safety
WARNING! Patient cable safety
To ensure defibrillation protection, you must use the Cardiac
Science-approved patient cable. Follow all instructions for the
patient cable.