Product Safety and Setup
Acquisition
Lead selection
Modes
Frequency
response
Input impedance
Electrode offset
tolerance
Sampling
Artifact filter
response
Storage
Pacemaker display
capability
Interpretation
(optional)
Environmental (Electrodes)
Operating and
storage
36
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 and Alternate
Chest Lead (chest lead selection V2R through V9R, V7, V8,
and V9)
Automatic, automatic rhythm, or manual rhythm
Meets or exceeds IEC 60601-2-51 standards
Meets or exceeds IEC 60601-2-51 standards
±300 mV
8000 Samples/Sec. 2.5μV LSB, Burst acquisition <100 μS
40 Hz, -3db
500 samples/second, 2.5μV LSB
Meets or exceeds IEC 60601-2-51
Diagnosis, measurements, reasons statements based on five
demographic criteria
Follow the electrode package labeling for operating and
storage. To prevent electrodes from drying out Cardiac
Science recommends storing electrodes in a reclosable
pouch.
70-01047-01 B