Table 3: Electromagnetic immunity (continued)
Immunity
IEC 60601 test
test
level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
c
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE 1: At 80 MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
70-01047-01 B
Compliance
level
3 Vrms
3 V/m
EMC guidelines
Electromagnetic
environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Burdick 8300/8500, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d
=
1.2 P
d
=
1.2 P
80 MHz to 800 MHz
d
=
2.3 P
800 MHz to 2,5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey,
should be less than the
compliance level in each frequency
b
.
range
Interference may occur in the vicinity of
equipment marked with the following
symbol:
43