Product Safety and Setup
WARNING! Trip hazard.
Route all cables away from main work areas to minimize risk of
tripping and injury.
WARNING! Shock hazard
If connecting the ECG device to non-medical equipment, use only
non-medical equipment compliant with IEC 60950 or 60601-1.
In addition, the enclosure leakage current of non-medical
equipment connected to the ECG device must not exceed 150 μA
in fault condition.
If necessary, use an additional isolating transformer or floating
power supply to maintain a proper enclosure leakage current and
provide additional protection.
WARNING! Misdiagnosis
Using unshielded or excessively long cables may cause or increase
susceptibility to electromagnetic interference.
Always use the shortest possible shielded cables.
Caution: Equipment damage
You must operate the ECG device must be operated only at the
line voltage and frequency specified on the external medical
grade power supply.
Grounding
All connected equipment must be IEC 60950 approved or
equivalent. Consult a qualified technician to verify equipment
compatibility.