ERROR STATEMENTS
ERROR
E. D. Un.
E. D. Ov.
E. T. Un. .
E. T. Ov.
E. I. Un.
E. I. Ov.
E. HW.
E. EE
E. . DATA
E. CLR.
|
76
INSTRUCTIONS FOR USE OM 402UNI
CAUSE
Number is too small (large negative) to be
displayed
Number is too large to be displayed
Number is outside the table range
Number is outside the table range
Input quantity is smaller than permitted input
quantity range
Input quantity is larger than permitted input
quantity range
A part of the instrument does not work properly
Data in EEPROM corrupted
Data in EEPROM outside the range
Memory was empty
(presetting carried out)
ELIMINATION
change DP setting, channel constant setting
change DP setting, channel constant setting
increase table values, change input setting (chan-
nel constant setting)
increase table values, change input setting (chan-
nel constant setting)
change input signal value or input (range) setting
change input signal value or input (range) setting
send the instrument for repair
perform restoration of manufacture setting, upon
repeated error statement send instrument for
repair
perform restoration of manufacture setting, upon
repeated error statement send instrument for
repair
upon repeated error statement send instrument for
repair, possible failure in calibration