4 digit programmable universal instrument. dc voltmeter/ammeter,process monitor, ohmmeter,thermometer for pt 100/500/1 000,thermometer for ni 1 000, thermometer for,thermocouples displays for lin ,potentiometers. (85 pages)
3 1/2 digit programmable. dc voltmeter/ammeter. ac voltmeter/ammeter. process monitor. ohmmeter/ thermometer for pt 100/500/1 000. thermometer for ni 1 000. thermometer for thermocouples. display inst. for lin. potentiometers (40 pages)
Page 1
G U A R A N T E E Y E A R S OMB 451UNI OMB 452UNI 4 DIGIT PROGRAMMABLE UNIVERSAL INSTRUMENT DC VOLTMETER/AMMETER PROCESS MONITOR OHMMETER THERMOMETER FOR PT 100/500/1 000 THERMOMETER FOR NI 1 000 THERMOMETER FOR THERMOCOUPLES...
Page 2
IEC 980: 1993, chapter 6 The instruments are applicable for unlimited use in agricultural and industrial areas. CONNECTION Supply of energy from the main line has to be isolated from the measuring leads. ORBIT MERRET, spol. s r.o. Vodnanska 675/30 198 00 Prague 9 Czech Republic Tel: +420 - 281 040 200...
INSTRUMENT DESCRIPTION DESCRIPTION The OMB 451/452 model series are programmable, three-color panel bargraphs with auxiliary display and adjustable LCD scale. The instruments are designed as dimensional replacement of the ZEPAKOMP instruments. Available are types UNI, PWR and UQC. Type OM 402UNI is a multifunction instrument with the option of configuration for 8 various input options, easily configurable in the instrument menu.
Page 5
OML cable to connect the instrument to PC. It is manufactured in version RS 232 and USB and is compatible with all ORBIT MERRET instruments. Another option for connection is with the aid of data output RS 232 or RS 485 (without the need of the OML cable).
INSTRUMENT CONNECTION The instrument supply leads should not be in proximity of the incoming low-potential signals. Contactors, motors with larger input power should not be in proximity of the instrument. The leads into the instrument input (measured quantity) should be in sufficient distance from all power leads and appliances.
Page 7
INSTRUMENT CONNECTION OMB 451UQC INPUT C2/Reset INPUT B2 INPUT A2 INPUT C1/Reset INPUT B1 INPUT A1 Excitation Option A INPUT - U INPUT - U RxD/L+ INPUT - U TxD/L- *GND - U/I0,5 GND - U *GND - I5 GND - I Analog output - I INPUT - I INPUT - I...
INSTRUMENT SETTING • For expert users • Complete instrument menu • Access is password protected • Possibility to arrange items of the „User“ menu • Tree menu structure • For trained users • Only items necessary for instrument setting • Access is password protected •...
Page 9
The operation program is freely accessible (www.orbit.merret.cz) and the only requirement is the purchase of OML cable to connect the instrument to PC. It is manufactured in version RS 232 and USB and is compatible with all ORBIT MERRET instruments.
INSTRUMENT SETTING Setting and controlling the instrument is performed by means of 5 control keys located on the front panel. With the aid of these keys it is possble to browse through the operation menu and to select and set required values. Indication of measured input (red/green/orange LED) LCD scale...
INSTRUMENT SETTING Setting the decimal point and the minus sign DECIMAL POINT Its selection in the menu, upon modification of the number to be adjusted it is performed by the short push key with transition beyond the highest decade, when the decimal point starts flashing. Positioning is performed by THE MINUS SIGN Setting the minus sign is performed by the key on higher decade.
light light SETTING SETTING “LIGHT” LIGHT Simple programming menu - contains only items necessary for instrument setting and is protected by optional number code • For capable users • Only items necessary for instrument setting • Access is password protected •...
Page 13
light light SETTING Access password 142. 8 PASSW. Type of instruments Measuring range TYPE MODE 4-20mA Selecting projection and connection CONNEC. 2-WIRE FORM. A 00000. o CONNEC. EXT. 1TC C. J . T EM. FORM. A 000000 MIN A MAX A FORM.
Page 14
light light SETTING 142. 8 Entering access password PASSW. for access into the menu Access into instrument PASSW. menu PAS > 0 PAS = 0 - access into menu is protected ny number - access into menu is unrestricted, after code releasing keys you automaticaly move to first item of the menu...
Page 15
light light SETTING INSTRUCTIONS FOR USE OMB 450UNI...
light light SETTING Type “DC” 60 mV 150 mV 300 mV 1200mV MODE Selection of the instrument MODE measuring range Menu Measuring range 60 mV ±60 mV 150 mV ±150 mV = 60 mV 300 mV ±300 mV 1200mV ±1,2 V 100 V ±100 V = 500 V*...
Page 17
light light SETTING Setting for maximum MAX A input signal - position of the DP does not affect display Setting display projection MAX A for maximum value of projection input signal - the DP is automatically shifted after the value is confirmed - range of the setting is -99999…999999 = 100 Projection for 150 mV >...
Page 18
light light SETTING Type “PM” • • • MODE 0-5mA 0-20mA 4-20mA 0-10 V 0-40 V Er. 4 Selection of the instrument MODE measuring range Menu Range 0-5mA 0…5 mA 0-20mA 0…20 mA 4-20mA 4…20 mA = 4 - 20 mA 0-2 V ±2 V 0-5 V...
Page 19
light light SETTING Setting for maximum MAX A input signal Setting display projection - position of the DP does not affect display MAX A for maximum value of projection input signal - the DP is automatically shifted after the value is confirmed - range of the setting is -99999…999999 = 100 Projection for 20 mA >...
Page 20
light light SETTING Type “OHM” 100 R 10 K 100 K MODE Selection of instrument Menu Measuring range MODE measuring range 100 R 0…100 Ϊ 0…1 kΪ 10 k 0…10 kΪ = 100 Ϊ 100 k 0…100 kΪ AUTO Autorange Example Range 0…10 kΩ...
Page 21
light light SETTING Setting for maximum MAX A input signal - position of the DP does not affect display Setting display projection MAX A for maximum value of projection input signal - the DP is automatically shifted after the value is confirmed - range of the setting is -99999…999999 = 100 Projection for 1 kΩ...
Page 23
light light SETTING FORM. A 000000 00000. o 0000. o o Setting projection of the = 00000.o FORM. A decimal point - positioning of the DP is set here in the measuring mode Example Projection of DP on display > 000000 * subsequent item on the menu depends on instrument equipment 00000.
Page 24
light light SETTING Type “RTD-Ni” 5. 0 - 1k 6. 2 - 1k 5. 0 -10k 6. 2 -10k MODE Selection of instrument MODE measuring range Menu Measuring range 5.0-1k Ni 1 000 (5 000 ppm/°C) 6.2-1k Ni 1 000 (6 180 ppm/°C) = Ni 1 000 - 5 000 ppm/°C 5.0-10k Ni 10 000 (5 000 ppm/°C)
Page 25
light light SETTING FORM. A 000000 00000. o 0000. o o Setting projection of the = 00000.o FORM. A decimal point - positioning of the DP is set here in the measuring mode Example Projection of DP on display > 000000 * subsequent item on the menu depends on instrument equipment 00000.
Page 26
light light SETTING Type “T/C” > T/C B T/C E T/C J T/C K MODE > T/C N T/C R T/C S T/C T T/C L Selection of the type of Menu Type of thermocouple MODE thermocouple T/C B T/C E - setting the input range depends on the T/C J measuring range ordered...
Page 27
light light SETTING Setting temperature of C. J . T EM. cold junction Setting temperature of cold = 23 C. J . T EM. junction - range 0…99°C with compensation box Setting temperature of cold junction > C.J. TEM. = 35 Example FORM.
Page 28
light light SETTING Type “DU” Setting for minimum MIN A input signal Setting display projection - position of the DP does not affect display MIN A for minimum value of projection input signal - the DP is automatically shifted after the value is confirmed - range of the setting is -99999…999999 Example...
Page 29
light light SETTING FORM. A 000000 00000. o 0000. o o 000. o oo 00. o ooo 0. o oooo FLOA. P. Setting projection of the = 0000.oo FORM. A decimal point - positioning of the DP is set here in the measuring mode Example Projection of DP on display >...
Page 30
light light SETTING Type “RTD-Cu” 428-50 428-0. 1 426-50 426-0. 1 MODE Selection of instrument MODE measuring range Menu Measuring range = Cu 50/4 280 ppm 5.0-1k Ni 1 000 (5 000 ppm/°C) 6.2-1k Ni 1 000 (6 180 ppm/°C) 5.0-10k Ni 10 000 (5 000 ppm/°C) 6.2-10k Ni 10 000 (6 180 ppm/°C) Example...
Page 31
light light SETTING FORM. A 000000 00000. o 0000. o o Setting projection of the = 00000.o FORM. A decimal point - positioning of the DP is set here in the measuring mode Example Projection of DP on display > 0000.oo * subsequent item on the menu depends on instrument equipment 00000.
light light SETTING Setting boundary for LIM. L. 1 limit 1 Setting boundary for - contingent modification of hysteresis or LIM. L. 1 limit 1 delay may be performed in “PROFI” menu - range of the setting is -99999…999999 = 20 - default “Hystresis”=0 “Delay”=0 Example Setting limit 1 >...
Page 33
light light SETTING Setting boundary for LIM. L. 3 limit 3 Setting boundary for - contingent modification of hysteresis or LIM. L. 3 limit 3 delay may be performed in “PROFI” menu - range of the setting is -99999…999999 = 60 - default “Hystresis”=0 “Delay”=0 Setting limit 3 >...
light light SETTING 0-20mA Er. 4-20 4-20mA 0-5mA 0-2 V 0-5 V 0-10 V TYP. A. O . Setting the type of analog TYP. A. O . output Menu Range Description 0-20mA 0…20 mA E. 4-20mA 4…20 mA with indication of error statement (<3,6 mA) 4-20mA 4…20 mA 0-5mA...
light light SETTING Assigning the display value to the end of the MAX. A. O . AO range Assigning the display = 100 MAX. A. O . value to the end of the AO range - range of the setting is -99999…999999 Example Display value for the end of the AO range >...
light light SETTING Setting for the maximum MAX. BG. input signal Bargraph projection = 100 MAX. BG. setting for the maximum value of the input signal - range of the setting is -99999…999999 Example Bargraph value at max. input > MAX. BG. = 120 DIVIS.
Page 37
light light SETTING VALUES Selection of numerical VALUES description on LCD scale = Yes - autonomous control of numerical description on LCD scale Description selection > VALUES = NO Example BACKL. BACKL. 100% LCD backlight level BACKL. selection = 100 % - LCD backlight level can be contolled to match the ambient lighting conditions - range of setting:...
light light SETTING COLOR GREEN ORANGE Selecting the bargraph COLOR colour = Green - this is to select the bargraph colour in the basic mode “Single bar” - for other modes it is necessary to switch into the “PROFI” menu Selecting the bargraph colour >...
light light SETTING RE. CAL. Prior to execution of any modifications you Restoration of manufacture RE. CAL. calibration will be asked to confirm your selection. (YES) - in the event of error calibration it is feasible to restore manufacture calibration. Example Restoration of manufacture setting >...
SETTING Type “DU” flashing legend C. MIN. Only for type “DU” Calibration of input range - C. MIN. the potentiometer traveller in initial position - prior confirming the flashing “YES” sign the potentiometer traveller has to be in given idle position Example Calibration of the beginning of the range >...
LANG. CZECH ENGL. Selection of language in LANG. instrument menu - selection of language version of the = ENGL. instrument menu Language selection - ENGLISH > LANG. = ENGL. ENGL. PAS. LI. Setting new access PAS. LI. password Setting new access - upon setting the password to “000”...
light light SETTING instrument type SW: number - version setting the input type IDENT. OMB 452UNI 65-001 Instrument SW version IDENT. - the display shows the type of instrument Desctription indication, SW number, SW version and type of instrument current input setting (Mode) SW: number - version - if SW version contains a letter in first the input type...
Page 43
light light SETTING INSTRUCTIONS FOR USE OMB 450UNI...
profi profi SETTING SETTING “PROFI” PROFI Complete programming menu • contains complete instrument menu and is protected by optional number code • designed for expert users • preset from manufacture is menu LIGHT • For expert users • Complete instrument menu •...
Page 45
profi profi SETTING INSTRUCTIONS FOR USE OMB 450UNI...
profi profi SETTING SETTING “PROFI” - INPUT The primary instrument parameters are set in this menu INPUTS CLEAR Resetting internal CLEAR CHANNE. CONFIG. values Selection of measuring OUTPUT. CONFIG. range and parameters SERVIC. EXT. IN. Setting date and time for option with RTC KEYS.
Page 48
profi profi SETTING Programming sche 142. 8 PASSW. Access password INPUTS CHANNE. CLEAR CHAN. A SET. A MIN A CL. TAR. CL. M. M . CL. MEM. CONFIG. READ/S 0. 1 0. 2 20. 0 40. 0 MAX A P. TAR. A TYPE DC PM DU Cu...
Page 49
profi profi SETTING eme PROFI MENU OUTPUT. SERVIC. MEMORY REWRIT. MENU NO YES LIGHT PROFI 00. 0 0. 0 0 START RESTOR. RE. CAL. STOP 00. 0 0. 0 0 PERIOD. 00. 0 0. 0 0 RE. SET. TYPE LIMITS LIM 1 INP.
Page 50
profi profi SETTING 6.1.2c Selection of measuring range Selection of instrument MODE measuring range INPUTS CLEAR READ/S 60mV 100 R Menu Measuring range 60mV ±60 mV CHANNE. CONFIG. TYPE 150mV 150mV ±150 mV 300mV ±300 mV OUTPUT. MODE 300mV 10 k 1200mV ±1,2 V Menu...
Page 51
profi profi SETTING 6.1.2d Selection of type of sensor connection RTD OHM Selection of type of CONNEC. sensor connection INPUTS CLEAR READ/S 2-WIRE RTD OHM CHANNE. CONFIG. TYPE 3-WIRE 2-wire connection 2-WIRE OUTPUT. MODE 4-WIRE 3-wire connection 3-WIRE SERVIC.
Page 52
profi profi SETTING 6.1.2e Setting temperature of cold junction Setting temperature of C. J . T EM. cold junction INPUTS CLEAR READ/S - range 0…99°C with compensation box CHANNE. CONFIG. TYPE = 23°C OUTPUT. MODE SERVIC. EXT. IN. CONNEC. KEYS. C.
profi profi SETTING 6.1.3 Setting the real time clock Setting the real time clock (RTC) INPUTS CLEAR TIME 000000 Time setting TIME CHANNE. CONFIG. DATE - format 23.59.59 OUTPUT. Date setting DATE SERVIC. EXT. IN. - format DD.MM.YY KEYS.
profi profi SETTING 6.1.4b Selection of function “HOLD” Selection of function M. HOLD “HOLD” INPUTS CLEAR EXT. 1 DISPL. “HOLD” locks only the DISPL. value displayed CHANNE. CONFIG. EXT. 2 DIS. + AV. “HOLD” locks the value DIS. + A. O . OUTPUT.
Page 55
profi profi SETTING 6.1.5b Optional accessory functions of the keys - Temporary projection Temporary projection of DOC. LE. selected item INPUTS CLEAR LEFT FN. LE. - “Temporary” projection of selected value is displayed for the time of keystroke CHANNE. CONFIG. DOWN DOC.
Page 56
profi profi SETTING 6.1.5c Optional accessory functions of the keys - Direct access to item Assigning access to MNU. LE. selected menu item INPUTS CLEAR LEFT FN. LE. LIM 1 Direct access to item LIM 1 “LIM 1” CHANNE. CONFIG.
Page 57
profi profi SETTING INSTRUCTIONS FOR USE OMB 450UNI...
profi profi SETTING SETTING “PROFI” - CHANNELS primary instrument parameters are set in this menu INPUTS CHAN. A Setting parameters of CHAN. A CHANNE. MAT. FN. measuring “Channel” Setting parameters of OUTPUT. MIN. M AX MAT. FN. mathematic functions SERVIC.
Page 59
profi profi SETTING 6.2.1c Digital filters Selection of digital MODE F. A filters INPUTS CHAN. A SET. A MODE F. A - at times it is useful for better user projection of data on display to modify it CHANNE. MAT. FN. FILTER CON.
Page 60
profi profi SETTING 6.2.1d Projection format - positioning of decimal point Selection of decimal FORM. A point INPUTS CHAN. A SET. A 000000 - the instrument allows for classic projection of a number with positioning of the DP CHANNE. MAT. FN. FILTER 00000.
Page 61
profi profi SETTING 6.2.1f Selection of storing data into instrument memory Selection of storing data LOG. A into instrument memory INPUTS CHAN. A SET. A SAVE A - by selection in this item you allow to register values into instrument memory CHANNE.
profi profi SETTING 6.2.2a Mathematic functions Selection of mathematic MATH. F. functions INPUTS CHAN. A MATH. F. Mathematic functions are off CHANNE. MAT. FN. CON. A MULTIN. Multinominal MULTIN. OUTPUT. MIN. M AX CON. B 1/MUL. SERVIC. CON. C EXPON.
Page 63
profi profi SETTING 6.2.2b Mathematic functions - decimal point Selection of decimal FORM. M. point INPUTS CHAN. A MATH. F. 000000 - the instrument allows for classic projection of a number with positioning of the DP CHANNE. MAT. FN. CON. A 00000.
profi profi SETTING 6.2.2d Mathematic functions - selection of storing data into instrument memory Selection of storing data LOG. M into instrument memory INPUTS CHAN. A MATH. F. SAVE M - by selection in this item you allow to register values into instrument memory CHANNE.
Page 65
profi profi SETTING INSTRUCTIONS FOR USE OMB 450UNI...
profi profi SETTING SETTING „PROFI“ - OUTPUTS In this menu it is possible to set parame ters of the instrument output signals INPUTS MEMORY Setting data logging into MEMORY CHANNE. LIMITS memory Setting type and OUTPUT. DATA LIMITS parameters of limits SERVIC.
Page 67
profi profi SETTING 6.3.1b Setting data logging into instrument memory - RTC Start of data logging into START instrument memory - time format HH.MM.SS INPUTS MEMORY REWRIT. Stop data logging into STOP CHANNE. LIMITS START instrument memory - time format HH.MM.SS OUTPUT.
profi profi SETTING z - delay “TIM. L.” Dose limit: MODE L. > “DOSE” axis x > “P” Period Switch-on relay TYPE > SWITCH-ON LED signalization Switch-off relay TYPE > SWITCH-OFF LED signalization Window limit: MODE L. > “FROM . .” axis x >...
Page 69
profi profi SETTING 6.3.2a Selection of input for limits evaluation Selection evaluation INP. L. 1 of limits INPUTS MEMORY LIM 1 INP. L. 1 - selection of value from which the limit will be evaluated CHANNE. LIMITS LIM 2 MOD. L. 1 CHAN.
Page 70
profi profi SETTING 6.3.2c Selection of type of output Selection of type of TYP. L. 1 output INPUTS MEMORY LIM 1 INP. L. 1 CLOSE. Output switches on when CLOSE. condition is met CHANNE. LIM. LIM 2 MOD. L. 1 OPEN ...
profi profi SETTING 6.3.3c Selection of data output protocol Selection of the type of PROT. analog output INPUTS MEMORY BAUD ASCII Data protocol ASCII ASCII CHANNE. LIMITS ADDR. M. BUS Data protocol M. BUS OUTPUT. DATA ADR. P. B . DIN MessBus SERVIC.
Page 73
profi profi SETTING 6.3.4b Selection of the type of analog output Selection of the type of TYP. A. O . analog output INPUTS MEMORY INP. A. O . 0-20mA Type - 0…20 mA 0-20mA CHANNE. LIMITA TYP. A. O . Er4-20 ...
profi profi SETTING 6.3.5a Selection of input for display projection Selection display PERM. projection INPUTS MEMORY PERM. CHAN. A - selection of value which will be shown on the instrument display CHANNE. LIMITS BRIGHT FIL. A Projection of values CHAN.
profi profi SETTING 6.3.6a Bargraph - Selection of input to be displayed Selection of value INP. BG. - bargraph INPUTS MEMORY INP. BG. - selecting the value which is displayed by the bargraph CHANNE. LIMITS MOD. BG. CHAN. A Bargraph is OFF OUTPUT.
Page 76
profi profi SETTING 6.3.6c Bargraph - Setting of projection range Setting of the bargraph BARGR. projection range INPUTS MEMORY INP. BG. Setting of the bargraph CHANNE. LIMITS MOD. BG. MIN. BG. projection range for the OUTPUT. DATA MIN. BG. minimum input signal - Setting range is -99999…999999 SERVIC.
Page 77
profi profi SETTING Selecting numeric VALUES description on LCD scale INPUTS MEMORY INP. BG. DIVIS. - autonomous control of numeric description on the LCD scale CHANNE. LIMITS MOD. BG. VALUES OUTPUT. DATA MIN. BG. FORM. S Numbers are disabled SERVIC.
Page 78
profi profi SETTING Selecting the LCD BACKL. backlight brightness INPUTS MEMORY INP. BG. DIVIS. - by adjusting the LCD backlight brightness we can match the ambient lighting CHANNE. LIMITS MOD. BG. VALUES conditions OUTPUT. DATA MIN. BG. FORM. S Backlight is off SERVIC.
Page 79
profi profi SETTING INSTRUCTIONS FOR USE OMB 450UNI...
Page 80
profi profi SETTING 6.3.6f Bargraph - Bargraph colour selection Selecting bargraph BAND. 0 colour INPUTS MEMORY INP. BG. BAND. 0 - item “COLORS” is displayed only in selected mode (“BARGR. > MOD. BG.”) CHANNE. LIMITS MOD. BG. B. L IM 1 GREEN “3 COL.”...
Page 81
profi profi SETTING 6.3.6h Bargraph - Selection of inverted projection Selection of inverted B. MOD. 1 projection “Band 0” INPUTS MEMORY INP. BG. BAND. 0 NORMAL - item “COLORS” is displayed only in selected mode (“BARGR. > MOD. BG.”) CHANNE. LIMITS MOD.
profi profi SETTING SETTING “PROFI” - SERVIS The instrument service functions are set in this menu INPUTS MENU Selection of menu type MENU CHANNE. RESTOR. LIGHT/PROFI Restore instrument OUTPUT. CALIB. RESTOR. manufacture setting and SERVIC. LANG. calibration Input range calibration N.
profi profi SETTING 6.4.2 Restoration of manufacture setting Restoration of RESTOR. manufacture setting INPUTS MENU RE. CAL. - in the event of error setting or calibration, manufacture setting may be restored. CHANNE. RESTOR. RE. SET. OUTPUT. CALIB. SAVE Restoration of RE.
profi profi SETTING 6.4.3 Calibration - Input range Input range CALIB. calibration INPUTS MENU C. MIN. - when “C. MIN” is displayed, move the potentiometer traveller to the required CHANNE. RESTOR. C. MAX. minimum position and confirm by „Enter“, calibration is confirmed by “YES“ OUTPUT.
profi profi SETTING 6.4.5 Setting new access password Setting new password N. PASS. for access to LIGHT and PROFI menu INPUTS MENU PAS. LI. - this option allows to change the numeric CHANNE. RESTOR. PAS. PR. code, which blocks the access into LIGHT and PROFI Menu.
user user SETTING SETTING ITEMS INTO “USER” MENU • USER menu is designed for users who need to change only several items of the setting without the option to change the primary instrument setting (e.g. repeated change of limit setting) •...
Page 87
user user SETTING Setting sequence of items in “USER” menu In compiling USER menu from active LIGHT menu the items (max. 10) may be assigned a sequence, in which they will be projected in the menu setting projection sequence Example: Into USER menu were selected these items: (keys ) >...
METHOD OF MEASURING THE CJC Instrument with input for temperature measurement with thermocouple allows to set two types of measurement of cold junction. Measuring thermocouple OM xxx T/C Reference thermocouple WITH REFERENCE THERMOCOUPLE – a reference thermocouple may be located in the same place as the measuring instrument or in place with stable temperature/compensation box –...
DATA PROTOCOL The instruments communicate via serial line RS232 or RS485. For communication they use the ASCII protocol. Communication runs in the following format: ASCII: 8 bit, no parity, one stop bit DIN MessBus: 7 bit, even parity, one stop bit The transfer rate is adjustable in the instrument menu.
ERROR STATEMENTS ERROR CAUSE ELIMINATION Number is too small (large negative) to be change DP setting, channel constant setting E. D. Un. displayed Number is too large to be displayed change DP setting, channel constant setting E. D. Ov. Number is outside the table range increase table values, change input setting (chan- E.
TABLE OF SIGNS The instrument allows to add two descriptive characters to the classic numeric formats (at the expense of the number of displayed places). The setting is performed by means of a shifted ASCII code. Upon modification the first two places display the entered characters and the last two places the code of the relevant symbol from 0 to 95.
TECHNICAL DATA INPUT Voltage of lin. pot. 2,5 VDC/6 mA range is adjustable min. potentiometer resistance is 500 Ohm ±60 mV >100 MOhm Input U ±150 mV >100 MOhm Input U PROJECTION ±300 mV >100 MOhm Input U Display: 999999, intensive red or green ±1200 mV >100 MOhm Input U...
Page 93
TECHNICAL DATA DATA OUTPUTS POWER SUPPLY Protocols: ASCII, DIN MessBus, MODBUS, PROBUS Options: 10…30 V AC/DC, 13,5 VA, isolated, Data format: 8 bit + no parity + 1 stop bit (ASCII) - fuse inside (T 4000 mA) 7 bit + even parity + 1 stop bit (MessBus) 80…250 V AC/DC, 13,5 VA, isolated Rate: 600…230 400 Baud...
INSTR. DIMENSIONS AND INSTALLATION OMB 451 Front view Panel cut 160 mm 150 mm 372. 4 ~C Side view 107 mm 9 mm Panel thickness: 0,5...20 mm OMB 452 Front view Panel cut 160 mm 150 mm 372. 4 Side view 107 mm 9 mm Panel thickness: 0,5...20 mm...
ORBIT MERRET, spol.s r.o. and that our company has taken all measures to ensure conformity of all products of the types referred-to hereunder, which are being brought out to the market, with technical documentation and requirements of the appurtenant Czech statutory orders.