Orbit Merret OMB 451UNI Manual

4 digit programmable universal instrument
Table of Contents

Advertisement

Quick Links

G U A R A N T E E
Y E A R S
OMB 451UNI
OMB 452UNI
4 DIGIT PROGRAMMABLE
UNIVERSAL INSTRUMENT
DC VOLTMETER/AMMETER
PROCESS MONITOR
OHMMETER
THERMOMETER FOR PT 100/500/1 000
THERMOMETER FOR NI 1 000
THERMOMETER FOR THERMOCOUPLES
DISPLAYS FOR LIN. POTENTIOMETERS

Advertisement

Table of Contents
loading

Summary of Contents for Orbit Merret OMB 451UNI

  • Page 1 G U A R A N T E E Y E A R S OMB 451UNI OMB 452UNI 4 DIGIT PROGRAMMABLE UNIVERSAL INSTRUMENT DC VOLTMETER/AMMETER PROCESS MONITOR OHMMETER THERMOMETER FOR PT 100/500/1 000 THERMOMETER FOR NI 1 000 THERMOMETER FOR THERMOCOUPLES...
  • Page 2 IEC 980: 1993, chapter 6 The instruments are applicable for unlimited use in agricultural and industrial areas. CONNECTION Supply of energy from the main line has to be isolated from the measuring leads. ORBIT MERRET, spol. s r.o. Vodnanska 675/30 198 00 Prague 9 Czech Republic Tel: +420 - 281 040 200...
  • Page 3: Table Of Contents

    CONTENTS Contents ..................3 Instrument description .
  • Page 4: Instrument Description

    INSTRUMENT DESCRIPTION DESCRIPTION The OMB 451/452 model series are programmable, three-color panel bargraphs with auxiliary display and adjustable LCD scale. The instruments are designed as dimensional replacement of the ZEPAKOMP instruments. Available are types UNI, PWR and UQC. Type OM 402UNI is a multifunction instrument with the option of configuration for 8 various input options, easily configurable in the instrument menu.
  • Page 5 OML cable to connect the instrument to PC. It is manufactured in version RS 232 and USB and is compatible with all ORBIT MERRET instruments. Another option for connection is with the aid of data output RS 232 or RS 485 (without the need of the OML cable).
  • Page 6: Instrument Connection

    INSTRUMENT CONNECTION The instrument supply leads should not be in proximity of the incoming low-potential signals. Contactors, motors with larger input power should not be in proximity of the instrument. The leads into the instrument input (measured quantity) should be in sufficient distance from all power leads and appliances.
  • Page 7 INSTRUMENT CONNECTION OMB 451UQC INPUT C2/Reset INPUT B2 INPUT A2 INPUT C1/Reset INPUT B1 INPUT A1 Excitation Option A INPUT - U INPUT - U RxD/L+ INPUT - U TxD/L- *GND - U/I0,5 GND - U *GND - I5 GND - I Analog output - I INPUT - I INPUT - I...
  • Page 8: Instrument Setting

    INSTRUMENT SETTING • For expert users • Complete instrument menu • Access is password protected • Possibility to arrange items of the „User“ menu • Tree menu structure • For trained users • Only items necessary for instrument setting • Access is password protected •...
  • Page 9 The operation program is freely accessible (www.orbit.merret.cz) and the only requirement is the purchase of OML cable to connect the instrument to PC. It is manufactured in version RS 232 and USB and is compatible with all ORBIT MERRET instruments.
  • Page 10: Symbols Used In The Instructions

    INSTRUMENT SETTING Setting and controlling the instrument is performed by means of 5 control keys located on the front panel. With the aid of these keys it is possble to browse through the operation menu and to select and set required values. Indication of measured input (red/green/orange LED) LCD scale...
  • Page 11: Control Keys Function

    INSTRUMENT SETTING Setting the decimal point and the minus sign DECIMAL POINT Its selection in the menu, upon modification of the number to be adjusted it is performed by the short push key with transition beyond the highest decade, when the decimal point starts flashing. Positioning is performed by THE MINUS SIGN Setting the minus sign is performed by the key on higher decade.
  • Page 12: Setting "Light" Menu

    light light SETTING SETTING “LIGHT” LIGHT Simple programming menu - contains only items necessary for instrument setting and is protected by optional number code • For capable users • Only items necessary for instrument setting • Access is password protected •...
  • Page 13 light light SETTING Access password 142. 8 PASSW. Type of instruments Measuring range TYPE MODE 4-20mA Selecting projection and connection CONNEC. 2-WIRE FORM. A 00000. o CONNEC. EXT. 1TC C. J . T EM. FORM. A 000000 MIN A MAX A FORM.
  • Page 14 light light SETTING 142. 8 Entering access password PASSW. for access into the menu Access into instrument PASSW. menu PAS > 0 PAS = 0 - access into menu is protected ny number - access into menu is unrestricted, after code releasing keys you automaticaly move to first item of the menu...
  • Page 15 light light SETTING INSTRUCTIONS FOR USE OMB 450UNI...
  • Page 16: Setting Input

    light light SETTING Type “DC” 60 mV 150 mV 300 mV 1200mV MODE Selection of the instrument MODE measuring range Menu Measuring range 60 mV ±60 mV 150 mV ±150 mV = 60 mV 300 mV ±300 mV 1200mV ±1,2 V 100 V ±100 V = 500 V*...
  • Page 17 light light SETTING Setting for maximum MAX A input signal - position of the DP does not affect display Setting display projection MAX A for maximum value of projection input signal - the DP is automatically shifted after the value is confirmed - range of the setting is -99999…999999 = 100 Projection for 150 mV >...
  • Page 18 light light SETTING Type “PM” • • • MODE 0-5mA 0-20mA 4-20mA 0-10 V 0-40 V Er. 4 Selection of the instrument MODE measuring range Menu Range 0-5mA 0…5 mA 0-20mA 0…20 mA 4-20mA 4…20 mA = 4 - 20 mA 0-2 V ±2 V 0-5 V...
  • Page 19 light light SETTING Setting for maximum MAX A input signal Setting display projection - position of the DP does not affect display MAX A for maximum value of projection input signal - the DP is automatically shifted after the value is confirmed - range of the setting is -99999…999999 = 100 Projection for 20 mA >...
  • Page 20 light light SETTING Type “OHM” 100 R 10 K 100 K MODE Selection of instrument Menu Measuring range MODE measuring range 100 R 0…100 Ϊ 0…1 kΪ 10 k 0…10 kΪ = 100 Ϊ 100 k 0…100 kΪ AUTO Autorange Example Range 0…10 kΩ...
  • Page 21 light light SETTING Setting for maximum MAX A input signal - position of the DP does not affect display Setting display projection MAX A for maximum value of projection input signal - the DP is automatically shifted after the value is confirmed - range of the setting is -99999…999999 = 100 Projection for 1 kΩ...
  • Page 22 light light SETTING Type “RTD-Pt” MODE EU-100 EU-500 EU-1k0 US-100 RU- 50 RU-100 Selection of instrument MODE measuring range Menu Measuring range EU-100 Pt 100 (3 850 ppm/°C) EU-500 Pt 500 (3 850 ppm/°C) = Pt 100 EU-1k0 Pt 1000 (3 850 ppm/°C) US-100 Pt 100 (3 920 ppm/°C) RU-50...
  • Page 23 light light SETTING FORM. A 000000 00000. o 0000. o o Setting projection of the = 00000.o FORM. A decimal point - positioning of the DP is set here in the measuring mode Example Projection of DP on display > 000000 * subsequent item on the menu depends on instrument equipment 00000.
  • Page 24 light light SETTING Type “RTD-Ni” 5. 0 - 1k 6. 2 - 1k 5. 0 -10k 6. 2 -10k MODE Selection of instrument MODE measuring range Menu Measuring range 5.0-1k Ni 1 000 (5 000 ppm/°C) 6.2-1k Ni 1 000 (6 180 ppm/°C) = Ni 1 000 - 5 000 ppm/°C 5.0-10k Ni 10 000 (5 000 ppm/°C)
  • Page 25 light light SETTING FORM. A 000000 00000. o 0000. o o Setting projection of the = 00000.o FORM. A decimal point - positioning of the DP is set here in the measuring mode Example Projection of DP on display > 000000 * subsequent item on the menu depends on instrument equipment 00000.
  • Page 26 light light SETTING Type “T/C” > T/C B T/C E T/C J T/C K MODE > T/C N T/C R T/C S T/C T T/C L Selection of the type of Menu Type of thermocouple MODE thermocouple T/C B T/C E - setting the input range depends on the T/C J measuring range ordered...
  • Page 27 light light SETTING Setting temperature of C. J . T EM. cold junction Setting temperature of cold = 23 C. J . T EM. junction - range 0…99°C with compensation box Setting temperature of cold junction > C.J. TEM. = 35 Example FORM.
  • Page 28 light light SETTING Type “DU” Setting for minimum MIN A input signal Setting display projection - position of the DP does not affect display MIN A for minimum value of projection input signal - the DP is automatically shifted after the value is confirmed - range of the setting is -99999…999999 Example...
  • Page 29 light light SETTING FORM. A 000000 00000. o 0000. o o 000. o oo 00. o ooo 0. o oooo FLOA. P. Setting projection of the = 0000.oo FORM. A decimal point - positioning of the DP is set here in the measuring mode Example Projection of DP on display >...
  • Page 30 light light SETTING Type “RTD-Cu” 428-50 428-0. 1 426-50 426-0. 1 MODE Selection of instrument MODE measuring range Menu Measuring range = Cu 50/4 280 ppm 5.0-1k Ni 1 000 (5 000 ppm/°C) 6.2-1k Ni 1 000 (6 180 ppm/°C) 5.0-10k Ni 10 000 (5 000 ppm/°C) 6.2-10k Ni 10 000 (6 180 ppm/°C) Example...
  • Page 31 light light SETTING FORM. A 000000 00000. o 0000. o o Setting projection of the = 00000.o FORM. A decimal point - positioning of the DP is set here in the measuring mode Example Projection of DP on display > 0000.oo * subsequent item on the menu depends on instrument equipment 00000.
  • Page 32: Setting Limits

    light light SETTING Setting boundary for LIM. L. 1 limit 1 Setting boundary for - contingent modification of hysteresis or LIM. L. 1 limit 1 delay may be performed in “PROFI” menu - range of the setting is -99999…999999 = 20 - default “Hystresis”=0 “Delay”=0 Example Setting limit 1 >...
  • Page 33 light light SETTING Setting boundary for LIM. L. 3 limit 3 Setting boundary for - contingent modification of hysteresis or LIM. L. 3 limit 3 delay may be performed in “PROFI” menu - range of the setting is -99999…999999 = 60 - default “Hystresis”=0 “Delay”=0 Setting limit 3 >...
  • Page 34: Setting Analog Output

    light light SETTING 0-20mA Er. 4-20 4-20mA 0-5mA 0-2 V 0-5 V 0-10 V TYP. A. O . Setting the type of analog TYP. A. O . output Menu Range Description 0-20mA 0…20 mA E. 4-20mA 4…20 mA with indication of error statement (<3,6 mA) 4-20mA 4…20 mA 0-5mA...
  • Page 35: Nastavení Rozsahu Bargrafu

    light light SETTING Assigning the display value to the end of the MAX. A. O . AO range Assigning the display = 100 MAX. A. O . value to the end of the AO range - range of the setting is -99999…999999 Example Display value for the end of the AO range >...
  • Page 36: Nastavení Lcd Stupnice

    light light SETTING Setting for the maximum MAX. BG. input signal Bargraph projection = 100 MAX. BG. setting for the maximum value of the input signal - range of the setting is -99999…999999 Example Bargraph value at max. input > MAX. BG. = 120 DIVIS.
  • Page 37 light light SETTING VALUES Selection of numerical VALUES description on LCD scale = Yes - autonomous control of numerical description on LCD scale Description selection > VALUES = NO Example BACKL. BACKL. 100% LCD backlight level BACKL. selection = 100 % - LCD backlight level can be contolled to match the ambient lighting conditions - range of setting:...
  • Page 38: Setting Of Bargrahp

    light light SETTING COLOR GREEN ORANGE Selecting the bargraph COLOR colour = Green - this is to select the bargraph colour in the basic mode “Single bar” - for other modes it is necessary to switch into the “PROFI” menu Selecting the bargraph colour >...
  • Page 39: Restoration Of Manufacture Setting

    light light SETTING RE. CAL. Prior to execution of any modifications you Restoration of manufacture RE. CAL. calibration will be asked to confirm your selection. (YES) - in the event of error calibration it is feasible to restore manufacture calibration. Example Restoration of manufacture setting >...
  • Page 40: Calibration - Input Range (Du)

    SETTING Type “DU” flashing legend C. MIN. Only for type “DU” Calibration of input range - C. MIN. the potentiometer traveller in initial position - prior confirming the flashing “YES” sign the potentiometer traveller has to be in given idle position Example Calibration of the beginning of the range >...
  • Page 41: Selection Of Instrument Menu Language Version

    LANG. CZECH ENGL. Selection of language in LANG. instrument menu - selection of language version of the = ENGL. instrument menu Language selection - ENGLISH > LANG. = ENGL. ENGL. PAS. LI. Setting new access PAS. LI. password Setting new access - upon setting the password to “000”...
  • Page 42: Instrument Identification

    light light SETTING instrument type SW: number - version setting the input type IDENT. OMB 452UNI 65-001 Instrument SW version IDENT. - the display shows the type of instrument Desctription indication, SW number, SW version and type of instrument current input setting (Mode) SW: number - version - if SW version contains a letter in first the input type...
  • Page 43 light light SETTING INSTRUCTIONS FOR USE OMB 450UNI...
  • Page 44: Setting "Profi" Menu

    profi profi SETTING SETTING “PROFI” PROFI Complete programming menu • contains complete instrument menu and is protected by optional number code • designed for expert users • preset from manufacture is menu LIGHT • For expert users • Complete instrument menu •...
  • Page 45 profi profi SETTING INSTRUCTIONS FOR USE OMB 450UNI...
  • Page 46: Resetting Internal Values

    profi profi SETTING SETTING “PROFI” - INPUT The primary instrument parameters are set in this menu INPUTS CLEAR   Resetting internal CLEAR CHANNE. CONFIG. values   Selection of measuring OUTPUT. CONFIG. range and parameters SERVIC. EXT. IN.   Setting date and time for option with RTC KEYS.
  • Page 47: Setting Measuring Type, Range, Mode, Rate

    profi profi SETTING 6.1.2a Selection of measuring rate Selection of measuring READ/S rate INPUTS CLEAR READ/S 40. 0   40,0 measurements/s 40. 0 CHANNE. CONFIG. TYPE 20. 0   20,0 measurements/s 20. 0 OUTPUT. MODE 10. 0   10,0 measurements/s SERVIC.
  • Page 48 profi profi SETTING Programming sche 142. 8 PASSW. Access password INPUTS CHANNE. CLEAR CHAN. A SET. A MIN A CL. TAR. CL. M. M . CL. MEM. CONFIG. READ/S 0. 1 0. 2 20. 0 40. 0 MAX A P. TAR. A TYPE DC PM DU Cu...
  • Page 49 profi profi SETTING eme PROFI MENU OUTPUT. SERVIC. MEMORY REWRIT. MENU NO YES LIGHT PROFI 00. 0 0. 0 0 START RESTOR. RE. CAL. STOP 00. 0 0. 0 0 PERIOD. 00. 0 0. 0 0 RE. SET. TYPE LIMITS LIM 1 INP.
  • Page 50 profi profi SETTING 6.1.2c Selection of measuring range Selection of instrument MODE measuring range INPUTS CLEAR READ/S 60mV 100 R Menu Measuring range 60mV ±60 mV CHANNE. CONFIG. TYPE 150mV 150mV ±150 mV 300mV ±300 mV OUTPUT. MODE 300mV 10 k 1200mV ±1,2 V Menu...
  • Page 51 profi profi SETTING 6.1.2d Selection of type of sensor connection RTD OHM Selection of type of CONNEC. sensor connection INPUTS CLEAR READ/S 2-WIRE RTD OHM   CHANNE. CONFIG. TYPE 3-WIRE 2-wire connection 2-WIRE OUTPUT. MODE 4-WIRE   3-wire connection 3-WIRE SERVIC.
  • Page 52 profi profi SETTING 6.1.2e Setting temperature of cold junction Setting temperature of C. J . T EM. cold junction INPUTS CLEAR READ/S - range 0…99°C with compensation box CHANNE. CONFIG. TYPE = 23°C OUTPUT. MODE SERVIC. EXT. IN. CONNEC. KEYS. C.
  • Page 53: Setting The Real Time

    profi profi SETTING 6.1.3 Setting the real time clock Setting the real time clock (RTC) INPUTS CLEAR TIME 000000   Time setting TIME CHANNE. CONFIG. DATE - format 23.59.59 OUTPUT.   Date setting DATE SERVIC. EXT. IN. - format DD.MM.YY KEYS.
  • Page 54: Optional Accessory Functions Of The Keys

    profi profi SETTING 6.1.4b Selection of function “HOLD” Selection of function M. HOLD “HOLD” INPUTS CLEAR EXT. 1 DISPL.   “HOLD” locks only the DISPL. value displayed CHANNE. CONFIG. EXT. 2 DIS. + AV.   “HOLD” locks the value DIS. + A. O . OUTPUT.
  • Page 55 profi profi SETTING 6.1.5b Optional accessory functions of the keys - Temporary projection Temporary projection of DOC. LE. selected item INPUTS CLEAR LEFT FN. LE. - “Temporary” projection of selected value is displayed for the time of keystroke CHANNE. CONFIG. DOWN DOC.
  • Page 56 profi profi SETTING 6.1.5c Optional accessory functions of the keys - Direct access to item Assigning access to MNU. LE. selected menu item INPUTS CLEAR LEFT FN. LE. LIM 1   Direct access to item LIM 1 “LIM 1” CHANNE. CONFIG.
  • Page 57 profi profi SETTING INSTRUCTIONS FOR USE OMB 450UNI...
  • Page 58: Setting Measuring Parameters (Projection, Filters, Decimal Point, Description)

    profi profi SETTING SETTING “PROFI” - CHANNELS primary instrument parameters are set in this menu INPUTS CHAN. A   Setting parameters of CHAN. A CHANNE. MAT. FN. measuring “Channel”   Setting parameters of OUTPUT. MIN. M AX MAT. FN. mathematic functions SERVIC.
  • Page 59 profi profi SETTING 6.2.1c Digital filters Selection of digital MODE F. A filters INPUTS CHAN. A SET. A MODE F. A - at times it is useful for better user projection of data on display to modify it CHANNE. MAT. FN. FILTER CON.
  • Page 60 profi profi SETTING 6.2.1d Projection format - positioning of decimal point Selection of decimal FORM. A point INPUTS CHAN. A SET. A 000000 - the instrument allows for classic projection of a number with positioning of the DP CHANNE. MAT. FN. FILTER 00000.
  • Page 61 profi profi SETTING 6.2.1f Selection of storing data into instrument memory Selection of storing data LOG. A into instrument memory INPUTS CHAN. A SET. A SAVE A - by selection in this item you allow to register values into instrument memory CHANNE.
  • Page 62: Setting Mathematic Functions

    profi profi SETTING 6.2.2a Mathematic functions Selection of mathematic MATH. F. functions INPUTS CHAN. A MATH. F.   Mathematic functions are off CHANNE. MAT. FN. CON. A MULTIN.   Multinominal MULTIN. OUTPUT. MIN. M AX CON. B 1/MUL. SERVIC. CON. C EXPON.
  • Page 63 profi profi SETTING 6.2.2b Mathematic functions - decimal point Selection of decimal FORM. M. point INPUTS CHAN. A MATH. F. 000000 - the instrument allows for classic projection of a number with positioning of the DP CHANNE. MAT. FN. CON. A 00000.
  • Page 64: Selection Of Evaluation Of Min/Max Value

    profi profi SETTING 6.2.2d Mathematic functions - selection of storing data into instrument memory Selection of storing data LOG. M into instrument memory INPUTS CHAN. A MATH. F. SAVE M - by selection in this item you allow to register values into instrument memory CHANNE.
  • Page 65 profi profi SETTING INSTRUCTIONS FOR USE OMB 450UNI...
  • Page 66: Setting Data Logging

    profi profi SETTING SETTING „PROFI“ - OUTPUTS In this menu it is possible to set parame ters of the instrument output signals INPUTS MEMORY   Setting data logging into MEMORY CHANNE. LIMITS memory   Setting type and OUTPUT. DATA LIMITS parameters of limits SERVIC.
  • Page 67 profi profi SETTING 6.3.1b Setting data logging into instrument memory - RTC   Start of data logging into START instrument memory - time format HH.MM.SS INPUTS MEMORY REWRIT.   Stop data logging into STOP CHANNE. LIMITS START instrument memory - time format HH.MM.SS OUTPUT.
  • Page 68: Setting Limits

    profi profi SETTING z - delay “TIM. L.” Dose limit: MODE L. > “DOSE” axis x > “P” Period Switch-on relay TYPE > SWITCH-ON LED signalization Switch-off relay TYPE > SWITCH-OFF LED signalization Window limit: MODE L. > “FROM . .” axis x >...
  • Page 69 profi profi SETTING 6.3.2a Selection of input for limits evaluation Selection evaluation INP. L. 1 of limits INPUTS MEMORY LIM 1 INP. L. 1 - selection of value from which the limit will be evaluated CHANNE. LIMITS LIM 2 MOD. L. 1 CHAN.
  • Page 70 profi profi SETTING 6.3.2c Selection of type of output Selection of type of TYP. L. 1 output INPUTS MEMORY LIM 1 INP. L. 1 CLOSE.   Output switches on when CLOSE. condition is met CHANNE. LIM. LIM 2 MOD. L. 1 OPEN  ...
  • Page 71: Setting Data Output

    profi profi SETTING 6.3.3a Selection of data output baud rate Selection of data output BAUD baud rate INPUTS MEMORY BAUD   Rate - 600 Baud CHANNE. LIMITS ADDR. 1200   Rate - 1 200 Baud 1200 OUTPUT. DATA ADR. P. B . 2400  ...
  • Page 72: Setting Analog Output

    profi profi SETTING 6.3.3c Selection of data output protocol Selection of the type of PROT. analog output INPUTS MEMORY BAUD ASCII   Data protocol ASCII ASCII CHANNE. LIMITS ADDR. M. BUS   Data protocol M. BUS OUTPUT. DATA ADR. P. B . DIN MessBus SERVIC.
  • Page 73 profi profi SETTING 6.3.4b Selection of the type of analog output Selection of the type of TYP. A. O . analog output INPUTS MEMORY INP. A. O . 0-20mA   Type - 0…20 mA 0-20mA CHANNE. LIMITA TYP. A. O . Er4-20  ...
  • Page 74: Selection Of Display Projection

    profi profi SETTING 6.3.5a Selection of input for display projection Selection display PERM. projection INPUTS MEMORY PERM. CHAN. A - selection of value which will be shown on the instrument display CHANNE. LIMITS BRIGHT FIL. A   Projection of values CHAN.
  • Page 75: Selection Of Bargraph Projection

    profi profi SETTING 6.3.6a Bargraph - Selection of input to be displayed Selection of value INP. BG. - bargraph INPUTS MEMORY INP. BG. - selecting the value which is displayed by the bargraph CHANNE. LIMITS MOD. BG. CHAN. A   Bargraph is OFF OUTPUT.
  • Page 76 profi profi SETTING 6.3.6c Bargraph - Setting of projection range Setting of the bargraph BARGR. projection range INPUTS MEMORY INP. BG.   Setting of the bargraph CHANNE. LIMITS MOD. BG. MIN. BG. projection range for the OUTPUT. DATA MIN. BG. minimum input signal - Setting range is -99999…999999 SERVIC.
  • Page 77 profi profi SETTING Selecting numeric VALUES description on LCD scale INPUTS MEMORY INP. BG. DIVIS. - autonomous control of numeric description on the LCD scale CHANNE. LIMITS MOD. BG. VALUES   OUTPUT. DATA MIN. BG. FORM. S Numbers are disabled SERVIC.
  • Page 78 profi profi SETTING Selecting the LCD BACKL. backlight brightness INPUTS MEMORY INP. BG. DIVIS. - by adjusting the LCD backlight brightness we can match the ambient lighting CHANNE. LIMITS MOD. BG. VALUES conditions OUTPUT. DATA MIN. BG. FORM. S   Backlight is off SERVIC.
  • Page 79 profi profi SETTING INSTRUCTIONS FOR USE OMB 450UNI...
  • Page 80 profi profi SETTING 6.3.6f Bargraph - Bargraph colour selection Selecting bargraph BAND. 0 colour INPUTS MEMORY INP. BG. BAND. 0 - item “COLORS” is displayed only in selected mode (“BARGR. > MOD. BG.”) CHANNE. LIMITS MOD. BG. B. L IM 1 GREEN “3 COL.”...
  • Page 81 profi profi SETTING 6.3.6h Bargraph - Selection of inverted projection Selection of inverted B. MOD. 1 projection “Band 0” INPUTS MEMORY INP. BG. BAND. 0 NORMAL - item “COLORS” is displayed only in selected mode (“BARGR. > MOD. BG.”) CHANNE. LIMITS MOD.
  • Page 82: Profi" Menu - Service 6.4.1 Selection Of Programming Menu „Light"/„Profi

    profi profi SETTING SETTING “PROFI” - SERVIS The instrument service functions are set in this menu INPUTS MENU   Selection of menu type MENU CHANNE. RESTOR. LIGHT/PROFI   Restore instrument OUTPUT. CALIB. RESTOR. manufacture setting and SERVIC. LANG. calibration   Input range calibration N.
  • Page 83: Restoration Manufacture Setting

    profi profi SETTING 6.4.2 Restoration of manufacture setting Restoration of RESTOR. manufacture setting INPUTS MENU RE. CAL. - in the event of error setting or calibration, manufacture setting may be restored. CHANNE. RESTOR. RE. SET. OUTPUT. CALIB. SAVE   Restoration of RE.
  • Page 84: Calibration - Input Range (Du)

    profi profi SETTING 6.4.3 Calibration - Input range Input range CALIB. calibration INPUTS MENU C. MIN. - when “C. MIN” is displayed, move the potentiometer traveller to the required CHANNE. RESTOR. C. MAX. minimum position and confirm by „Enter“, calibration is confirmed by “YES“ OUTPUT.
  • Page 85: Setting New Access Password

    profi profi SETTING 6.4.5 Setting new access password Setting new password N. PASS. for access to LIGHT and PROFI menu INPUTS MENU PAS. LI. - this option allows to change the numeric CHANNE. RESTOR. PAS. PR. code, which blocks the access into LIGHT and PROFI Menu.
  • Page 86: Setting Items Into "User" Menu

    user user SETTING SETTING ITEMS INTO “USER” MENU • USER menu is designed for users who need to change only several items of the setting without the option to change the primary instrument setting (e.g. repeated change of limit setting) •...
  • Page 87 user user SETTING Setting sequence of items in “USER” menu In compiling USER menu from active LIGHT menu the items (max. 10) may be assigned a sequence, in which they will be projected in the menu setting projection sequence Example: Into USER menu were selected these items: (keys ) >...
  • Page 88: Method Of Measuring Of The Cold Junction

    METHOD OF MEASURING THE CJC Instrument with input for temperature measurement with thermocouple allows to set two types of measurement of cold junction. Measuring thermocouple OM xxx T/C Reference thermocouple WITH REFERENCE THERMOCOUPLE – a reference thermocouple may be located in the same place as the measuring instrument or in place with stable temperature/compensation box –...
  • Page 89: Data Protocol

    DATA PROTOCOL The instruments communicate via serial line RS232 or RS485. For communication they use the ASCII protocol. Communication runs in the following format: ASCII: 8 bit, no parity, one stop bit DIN MessBus: 7 bit, even parity, one stop bit The transfer rate is adjustable in the instrument menu.
  • Page 90: Error Statements

    ERROR STATEMENTS ERROR CAUSE ELIMINATION Number is too small (large negative) to be change DP setting, channel constant setting E. D. Un. displayed Number is too large to be displayed change DP setting, channel constant setting E. D. Ov. Number is outside the table range increase table values, change input setting (chan- E.
  • Page 91: Table Of Symbols

    TABLE OF SIGNS The instrument allows to add two descriptive characters to the classic numeric formats (at the expense of the number of displayed places). The setting is performed by means of a shifted ASCII code. Upon modification the first two places display the entered characters and the last two places the code of the relevant symbol from 0 to 95.
  • Page 92: Technical Data

    TECHNICAL DATA INPUT Voltage of lin. pot. 2,5 VDC/6 mA range is adjustable min. potentiometer resistance is 500 Ohm ±60 mV >100 MOhm Input U ±150 mV >100 MOhm Input U PROJECTION ±300 mV >100 MOhm Input U Display: 999999, intensive red or green ±1200 mV >100 MOhm Input U...
  • Page 93 TECHNICAL DATA DATA OUTPUTS POWER SUPPLY Protocols: ASCII, DIN MessBus, MODBUS, PROBUS Options: 10…30 V AC/DC, 13,5 VA, isolated, Data format: 8 bit + no parity + 1 stop bit (ASCII) - fuse inside (T 4000 mA) 7 bit + even parity + 1 stop bit (MessBus) 80…250 V AC/DC, 13,5 VA, isolated Rate: 600…230 400 Baud...
  • Page 94: Instrument Dimensions And Instalation

    INSTR. DIMENSIONS AND INSTALLATION OMB 451 Front view Panel cut 160 mm 150 mm 372. 4 ~C Side view 107 mm 9 mm Panel thickness: 0,5...20 mm OMB 452 Front view Panel cut 160 mm 150 mm 372. 4 Side view 107 mm 9 mm Panel thickness: 0,5...20 mm...
  • Page 95: Certificate Of Guarantee

    CERTIFICATE OF GUARANTEE Product OMB 451UNI OMB 452UNI Type ....Manufacturing No.....
  • Page 96: Declaration Of Conformity

    ORBIT MERRET, spol.s r.o. and that our company has taken all measures to ensure conformity of all products of the types referred-to hereunder, which are being brought out to the market, with technical documentation and requirements of the appurtenant Czech statutory orders.

This manual is also suitable for:

Omb 452uni

Table of Contents