HP E3473A User Manual page 46

For sh7040/50
Table of Contents

Advertisement

Switch S5 should be set to OPEN if you are connecting to the 10BASE-T
connector.
Set all other switches to CLOSED.
Connect the power cord to the HP E3472A/73A Emulator.
5
Verify that the power light stays on after 10 seconds.
The IP address will be stored in EEPROM.
Disconnect the power cable from the emulation controller. Set
6
switch S6 back to CLOSED and connect the power cable again.
Do this so that the HP E3472A/73A Emulator does not request its IP address
each time power is cycled. The IP address is stored in EEPROM, so BOOTP
does not need to be run again. Leaving this switch on will result in slower
performance, increased LAN traffic, and even failure to power up (if the
BOOTP server becomes inactive).
Verify your HP E3472A/73A Emulator is now active and on the
7
network. See "To verify LAN communications" in this chapter.
Connecting to the Host Computer
To configure LAN parameters using BOOTP
31