9 Appendix A – Technical data
QIAGEN reserves the right to change specifications at any time.
9.1 Operating conditions
Power
100–240 V AC, 50/60 Hz, 650 VA Mains supply voltage
fluctuations are not to exceed 10% of nominal supply voltages.
Note: The apparent power can exceed 650 VA for up to 2 seconds
during the centrifuge acceleration and can reach an approximate
value of 1000 VA.
Fuse
2x T6.3A L 250V
Overvoltage
II
category
Air temperature
18 to 28ºC (64.4 to 82.4ºF)
Relative humidity
15–75% (noncondensing)
Altitude
Up to 2000 m (6500 ft.)
Place of operation
For indoor use only
Pollution level
2
Environmental class
3K2 (IEC 60721-3-3)
9.2 Transport conditions
Air temperature
–25ºC to 60ºC (–13ºF to 140ºF) in manufacturer's package
Relative humidity
Max. 75% (noncondensing)
Environmental class
2K2 & 2M2 (IEC 60721-3-2)
9.3 Storage conditions
Air temperature
15ºC to 30ºC (59ºF to 86ºF) in manufacturer's package
Relative humidity
Max. 75% (noncondensing)
Environmental class
1K2 (IEC 60721-3-1)
QIAcube Connect User Manual 09/2019
133