Replacing The Glow Plug - Teledyne FALCON GC Operation And Maintenance Manual

Table of Contents

Advertisement

Falcon GC
20. Replace the front on the unit using the 4 screws removed in Step
1.
Installation is complete. After powering up the instrument, all of the
column's information can be viewed either from the touchscreen display
in the "Column Module" screen or in under the "GC Info" tab of the
"Special Features" box in Chromperfect.

5.6 Replacing the Glow Plug

Caution:
If the unit is on and at operating temperature the
glow plug will be very hot. It should NOT be
touched without proper precautions. If the unit is
off and near room temperature the glow plug will
be at a safe temperature and can be touched
without tools or gloves.
1. Remove the red wire that connects the glow plug to the FID
board near the base of the FID. This can be done by lightly
pulling on the rubber sleeve covering the metal end of the glow
plug wire.
Teledyne Analytical Instruments
Maintenance & Troubleshooting